Owner manual
5
LS-Navigator-M-NO5-E
5-22
5-10
Setting the Measurement Conditions for the Program
This section describes how to set the measurement conditions for [Limits], [Calibration], [Area 
settings], and [Output settings].
1 
Click on the [Limits] tab in the program 
setting area.
The [Limits] window appears.
2 
Set [Limits] for OUT1 and OUT2.
Note
・ Use the controller to change settings in limit mode.
For more details, see the “Tolerance Settings” section in the “User’s Manual” for each controller.
・ Set the same values for [Units] and [Minimum display unit] in the settings for the controller and LS-Navigator. (Refer to 
Page 5-5.)
Limits
Select the output (OUT1 
or OUT2).
Set the standard value, as well as the 
upper and lower limits for the deviation.
The measurement is judged as one of 
three levels: HI, GO, or LO.
Set the upper limits and lower limits with 
absolute values. The measurement is judged 
as one of five levels: HH, HI, GO, LO, or LL.
Reference
The range for each parameter depends on the 
settings for [Units] and [Minimum display unit].
If dust or foreign objects become attached to 
the measurement target, the displayed 
measurement value may become extremely 
inaccurate. If dust or foreign objects become 
attached to the measurement target, the 
displayed measurement value may become 
extremely inaccurate
. When this para
meter is 
set on, measurement values that exceed the 
upper limit for errors (HOLD-H) or fall below the 
lower limit for errors (HOLD-L) are automatically 
detected and eliminated. Select [ON] from the 
pull-down menu and enter the upper limit or 
lower limit for errors in the text box.
Copies the limit settings from 
OUT1 or OUT2 and pastes 
them into OUT1 or OUT2.
1










