96M12221 User’s Manual High-speed, High-precision Digital Micrometer LS-7600 Series
Preface This instruction manual provides necessary information on the operation and maintenance of the LS-7600 Series along with precautions. Please read the manual carefully and be sure you understand the information provided before attempting to install and operate the LS-7600 Series. Keep this manual handy for future reference. Please make sure that the end users are provided with this manual.
Safety Precautions ■ General precautions • At startup and during operation, be sure to monitor the functions and performance of the LS-7600 Series. • Take sufficient safety measures to prevent damage to the human body and/or equipment that may be caused if this product should fail to operate properly. • Do not modify the LS-7600 Series or use it in any way other than as described in the specifications. Its functions and performance are not guaranteed under said conditions.
CAUTION ■ Usage • Be sure to turn off the power to the LS-7600 Series and any connected devices before connecting or disconnecting the cables. Otherwise, the camera and connected devices may be damaged. • Do not turn off the power while setting a parameter. Otherwise, the settings may be partially or completely lost. • Do not block the ventilation slots on the LS-7600 Series and peripheral devices. A rise in inner temperature may cause equipment failure.
Note: ■ Influence of Ambient Operating Temperature Ensure that the ambient operating temperature is constant. A change in the ambient operating temperature may result in measurement errors. If the ambient operating temperature changes by 10°C, it will take approximately 60 minutes for the interior temperature distribution of the LS-7600 Series to be uniform. ■ Warming Up Do not operate the LS-7600 Series for approximately 30 minutes after it is turned on.
Precautions for CE Markings Keyence confirms that the LS-7600 Series meets EC directive requirements. The LS-7600 Series bears CE markings. Keep the following conditions if the LS-7600 Series is used in European countries. ■ EMC Directives Keep the following condition so that the LS-7600 Series will satisfy EN61326-1 requirements. • The power cable and I/O cable connected to the controller are both less than 30m in length.
About this Manual The following section provides information on the configuration of each page of this manual along with symbols and terms used in this manual. Page Configuration and Symbols Chapter 4 Operation Control 4-4 Automatic Zero Function The automatic zero function sets measuring values to zero (0.00000) instantly. This function is convenient for zero-point calibration for a variety of target.
Optimum Measurement Methods Refer to the following methods, and select the optimum measurement method according to the type of target.
Configuration of Manual Chapter 1 Getting Started 1 Provides information on precautions and necessary preparations required. Chapter 2 Basic Operation of LS-7600 Series 2 Provides information on the basic operation of the LS-7600 Series, panel keys, and remote control console. Chapter 3 Easy Setting Guide 3 Provides information on targets and settings for typical applications respectively.
Contents Safety Precautions .............................................................................................................. i Precautions for CE Markings ........................................................................................... iv About this Manual .............................................................................................................. v Page Configuration and Symbols ....................................................................................
Target Viewer Function ................................................................................................... 4-7 Switching Cameras ...................................................................................................... 4-8 Hold Function .................................................................................................................. 4-9 Automatic Zero Function ..............................................................................................
Chapter 6 Environment Settings Possible Environment Settings ...................................................................................... 6-2 Default Settings ............................................................................................................ 6-2 Changing Environment Settings .................................................................................. 6-4 Details of Environment Settings .................................................................................
Appendix List of Optional Accessories ........................................................................... Setting Record Sheet ....................................................................................... Environment Setting Record Sheet ................................................................ Index .................................................................................................................
MEMO xii
Chapter 1 Getting Started This section provides information on the configuration of the LS-7600 Series, precautions, and necessary preparations required before operating the LS-7600 Series. Familiarize yourself with this section before using the LS-7600 Series. Outline and Features of LS-7600 Series .............................. 1-2 System Applications ............................................................. 1-4 Package Contents ................................................................
Chapter 1 Getting Started Outline and Features of LS-7600 Series The LS-7600 Series is a high-speed, high-accuracy digital micrometer used for the dimensional measurement of objects without coming into contact with the objects. It is such a versatile model that it has a wide range of applications including in-line measurement and offline measurement.
Chapter 1 Getting Started ■ Principle of Measurement The high-intensity GaN green LED radiates light, which will be changed into uniform parallel light through the special diffusion unit and collimator lens and emitted to the target in the measuring range. Then the shadow image of the target will appear on the HL-CCD (high-speed linear CCD) through the telecentric optical system.
Chapter 1 Getting Started System Applications The LS-7600 Series has a wide range of applications when used in combination with commercially available peripheral devices. 1 PC Control values or measuring values are retrieved over RS232C. Indicator Light and Buzzer Alarms are generated according to the output of discrimination results. LS-7601 Photoelectric Sensor or Proximity Sensor Timing input signals are sent whenever targets are detected.
Chapter 1 Getting Started Package Contents Check that nothing is missing from the LS-7600 Series package before use.
Chapter 1 Getting Started LS-7070M Measuring Head 1 Allen-head bolt (Six, M4 x 50 with a washer) Allen-head bolt (Four, M5 x 20) LS-C✱✱A Extension Cable (Cable between the controller and measuring head) LS-C3A: 3-m cable ✱✱... LS-C10A: 10-m cable ✱✱... LS-C30A: 30-m cable Up to two cables are connectable, provided that the total length of the cables does not exceed 40 m. LS-C✱✱AM Camera cable LS-C3AM: 3-m cable ✱✱... LS-C10AM: 10-m cable ✱✱...
Chapter 1 Getting Started LS-S11 Stand Unit Base Unit 1 Screws (Four, M4 x 8) ❈ Keyence ships each package with utmost care and attention. If there should be any improper or damaged product, contact your Keyence representative.
Chapter 1 Getting Started Functions and Nomenclature The following section provides information on the functions and nomenclature of the controller and measuring head. 1 Controller LCD Monitor Control Panel (Page 2-2, Section 2 Basic Operation of LS-7600 Series) Power Supply Indicator The indicator is lit when power is supplied to the LCD monitor.
Chapter 1 Getting Started Measuring Head LS-7010M Multi-purpose Mounting Holes A fixture for targets can be attached. 1 Cover Glass Receiver (R Head) Receives light from the transmitter. Transmitter (T Head) Emits light for measurement. TR Base Used to mount the transmitter and receiver. TR Cables Used to connect the transmitter Mounting Holes Used to mount the measuring head. Extension Cable Connected to the controller. Camera Cable Connected to the controller.
Chapter 1 Getting Started LS-7070M Multi-purpose Mounting Holes A fixture for targets can be attached. 1 Cover Glass Receiver (R Head) Receives light from the transmitter. Transmitter (T Head) Emits light for measurement. TR Base Used to mount the transmitter and receiver. TR Cables Used to connect the transmitter and receiver. Mounting holes Used to mount the measuring head. Extension Cable Connected to the controller. Camera Cable Connected to the controller.
Chapter 1 Getting Started Mounting and Connection Mount the measuring head and the controller, and connect them with the cables. 1 ■ Mounting the Measuring Head The measuring head can be mounted with or without the TR base employed, according to the type of targets and operating environment. The following section provides information on how to mount the measuring head with or without the TR base employed, along with mounting conditions required.
Chapter 1 Getting Started ■ With TR Base The transmitter and receiver are mounted to the TR base at the time of shipment. This section provides how to install the transmitter and receiver (LS-7030M in this case) that are mounted on the TR base. LS-7010M and LS-7070M can also be installed in the same way. 1 Use the four, M4x20 Allen-head bolts, which are provided with the package, to mount the TR base.
Chapter 1 Getting Started Use the 5-mm-deep M4 bottom mounting holes of the transmitter and receiver as shown in the illustration. 1 Top ■ When Using the Dual Head Function When using the dual head function, mount the two measuring heads so that their top sides face each other. Mounting the Controller ■ Mounting Restrictions Be aware of the following items when mounting the controller. CAUTION Note: • Do not install the controller upside down. • Do not block the ventilation louver.
Chapter 1 Getting Started ■ Mounting to Panel The controller can be mounted to control panels with the mounting brackets provided with the package. Arrange the following panel mounting dimensions. 1 207 169 124 +1 0 +1 0 200 Unit: mm 1 Mount the controller from the front side of the panel. 2 Attach resin caps to the mounting brackets. 3 Insert the mounting brackets to the four mounting holes on the side of the controller respectively.
Chapter 1 Getting Started 4 Secure the mounting brackets with a Phillips screwdriver. 1 CAUTION • Tighten the mounting screws to a maximum torque of 0.4 Nm. Do not tighten up the mounting screws excessively, otherwise the panel or controller casing may be deformed. • Do not remove the gasket, otherwise the controller will not satisfy IP64 conditions.
Chapter 1 Getting Started 3 Connect the cable between the controller and measuring head. If only a single measuring head is used, connect the cable to the HEAD1 connector. 1 Cable between controller and measuring head 4 Connect the camera cable to the controller. If only a single measuring head is used, connect the cable to the CAMERA1 connector. Camera cable 5 Connect the power supply cable of the AC power supply stand to the controller. 6 Mount the stand unit to the base unit.
Chapter 1 Getting Started Tips The bottom of the AC power supply stand has legs. Erect the legs and use the AC power supply stand if required. 1 Connection Connect the measuring head, controller, remote control console, and power supply with the cables. Do not supply power to the controller while connecting the cables. If the AC power supply stand is used, turn off the AC power supply stand while connecting the cables.
Chapter 1 Getting Started ■ Transmitter and Receiver Connect the cables of the transmitter and receiver respectively. Connect both connectors so that the notched part of each connector will coincide with each other in position. 1 Transmitter side Receiver side ■ Measuring Head and Controller Connect the cables of the controller and the measuring head. Connect both connectors so that the notched part of each connector will coincide with each other in position.
Chapter 1 Getting Started Turning AC Power Supply Stand On and Off This section provides information on how to turn the AC power supply stand on and off. 1 ■ Turning Power On 1 Turn on the POWER switch on the rear panel of the AC power supply stand. ■ Turning Power Off 1 Turn off the POWER switch on the rear panel of the AC power supply stand.
Chapter 1 Getting Started Adjustments to LCD Monitor The color, hue, and brightness adjustments to the LCD monitor. The LCD monitor is factory-adjusted. Therefore, make these adjustments only if they are required according to the operating conditions. 1 Tips The LCD monitor will be reset to the initial state by setting each trimmer to the center position. ■ Color Adjustments Turn the COLOR adjustment trimmer.
Chapter 1 Getting Started Measurement and Setting Overview The LS-7600 Series has two modes (i.e., the RUN mode and PROGRAM mode). Program settings (measuring condition settings) and environment settings (basic settings) can be made in the LS-7601 in PROGRAM mode. Power On 1 RUN mode Measures according to the set program. Refer to Section 4 Operation Control on page 4-1. RUN PROG RUN PROG or the [Slide switch] key operation on the console.
Chapter 1 Getting Started Resetting to Initial Status The following section provides information on how to reset all the programs and environment settings to factory-set values. 1 Tips Refer to Resetting All Programs on page 6-6 if the resetting of only the programs to factory-set values is required. 1 Turn on the LS-7600 Series with the [Zero] key pressed. A confirmation message for resetting will appear on the LCD monitor screen. 2 Press the [Left or Right] key and select Y.
Chapter 2 Basic Operation of the LS-7600 Series This section provides information on the basic operation of the LS-7600 Series, panel keys, and remote control console. Control Panel Functions and Nomenclature ........................ 2-2 Names and Functions of the LCD Monitor Screen ...............
Chapter 2 Basic Operation of the LS-7600 Series Control Panel Functions and Nomenclature The LS-7600 Series is operated with the panel keys and remote control console. The LCD monitor will display the result of measurement and necessary data for setting. Panel Key Functions 2 [SCREEN] key Press this key to change the screen display. (Page 4-2, Section 4-1 Switching Display Screen) [RUN/PROG] key Press this key to change the LS7600 Series between RUN mode and PROG mode.
Chapter 2 Basic Operation of the LS-7600 Series Remote Control Console Functions and Nomenclature 2 [ENT] key Use this key to move the cursor or input the set value change. [ESCAPE] key Press this key to quit or abort the set item. When in the Run mode, this key has the same function as the RESET input terminal on the rear panel. RUN ENT PROG ESCAPE SCREEN [SCREEN] key Press this key to change the screen display.
Chapter 2 Basic Operation of the LS-7600 Series Names and Functions of the LCD Monitor Screen This section describes the contents displayed on the LCD monitor screen. For the operation method, refer to Switching Display Screen on page 4-2. ■ Single Display 2 Application mode display OUT number display Displays the current OUT number. Status display Target viewer display Displays the image of the measuring target. Target Viewer Function on page 4-7 Tolerance display Displays the preset tolerance values.
Chapter 3 Easy Setting Guide The LS-7600 Series is used for a variety of measurement applications if the LS-7600 Series is set properly. This section provides information on targets and settings for typical applications respectively. Measurement of Outer Diameter and Width (with Single Measuring Head) ............................................. 3-2 Measurement of Outer Diameter and Width (with Two Measuring Heads) ...............................................
Chapter 3 Easy Setting Guide Measurement of Outer Diameter and Width (with Single Measuring Head) In the following example, a single measuring head is used to measure the outer diameter of an object in one direction. The result of measurement is displayed as OUT1. D 3 Measures the outer diameter in one direction. ■ Set Values Set item Area setting Set value AREA 1 AREA Calibration Head 1 DIA Make settings if required. OUT 1 Operation Output setting Select the value according to the line speed.
Chapter 3 Easy Setting Guide Measurement of Outer Diameter and Width (with Two Measuring Heads) In the following example, two measuring heads are used to measure the average outer diameter of an object in two directions. The measurement of the object in two directions ensures high accuracy. The result of measurement is displayed as OUT1. 3 Y X D= X+Y 2 Measures the outer diameter in two directions.
Chapter 3 Easy Setting Guide Measurement of Outer Diameter and Width (with Two Measuring Heads for Larger Objects) A sheet cannot be located in the measuring area of a single measuring head if the width or diameter of the sheet is excessively large. In that case, the sheet can be measured with two measuring heads as shown below with the automatic zero function and the master workpiece used. The result of output is displayed as OUT1.
Chapter 3 Easy Setting Guide Measurement of Inner Diameter and Clearance In the following example, the inner diameter of the workpiece is measured while the workpiece moves. The maximum value measured is taken as the inner diameter of the workpiece. The result of output is displayed as OUT1. 3 ■ Set Values Set item Set value AREA 1 Area setting AREA Calibration Head 1 SEG (+1E, +2E) No settings are required. Make settings if required.
Chapter 3 Easy Setting Guide Measurement of Outer Diameter and Eccentricity In the following example, the outer diameter and eccentricity of the roller is measured while the roller moves. The outer diameter is displayed as OUT1, and the eccentricity is displayed as OUT2. 3 Eccentricity Outer diameter ■ Set Values Set item Area setting Set value AREA 2 AREA 1 AREA Calibration Head 1 DIA Head 1 T-EDGE Make settings if required.
Chapter 3 Easy Setting Guide Measurement of Movement and Positioning In the following example, the movement or position of the workpiece is measured. The result of output is displayed as OUT1. 3 ■ Set Values Set item Set value AREA 1 Area setting AREA Calibration (0E, +2E) No settings are required. Make settings if required. OUT 1 Output setting Operation A2: 0 Select the value according to the line speed.
Chapter 3 Easy Setting Guide Measurement of Pitch The single pitch of connectors pins or IC pins is measured by the LS-7600 Series. Furthermore, multiple pitch of pins is measured with the program function or RS-232C interface used. 3 The center pitches of the two portions are measured at a time. ■ Set Values Set item Set value AREA 2 AREA 1 Area setting AREA Calibration Head 1 SEG Head 1 SEG (+2P, +4P) (+4P, +6P) Make settings if required.
Chapter 3 Easy Setting Guide Group Processing In the following example, the average, maximum, or minimum value is calculated as criteria from the measuring data on the group of targets. The results of output are displayed as OUT2. Other example 3 The average measured value of the 12 shafts is The average measured value of portions used as criteria. A through D is used as criteria. ■ Set Values Set item Area setting Set value AREA 1 AREA Calibration Head 1 DIA Make settings if required.
Chapter 3 Easy Setting Guide Statistical Processing In the following example, the values calculated from the measuring data are: Maximum value, minimum value, average value, maximum and minimum values, standard deviation, N number, HH number, HI number, GO number, LO number, and LL number. 3 The outer diameters of 96 shafts sampled are measured to calculate the values. ■ Set Values Set item Area setting Set value AREA 1 AREA Calibration Head 1 DIA Make settings if required.
Chapter 4 Operation Control This section provides information on setting control items on the control panel while the LS-7600 Series is in measurement operation. Switching Display Screen ............................................ 4-2 Target Viewer Function ............................................... 4-7 Hold Function .............................................................. 4-9 Automatic Zero Function ............................................ 4-10 Key Lock Function ...........................
Chapter 4 Operation Control Switching Display Screen The following section provides information on how to display an OUT1 or OUT2 measured value or both on the LCD monitor. Furthermore, it provides information on trend graph, statistical processing and NG RECORD display. The LCD monitor is operable through the panel keys as well as the remote control console. Measured Value Display The OUT1 or OUT2 measured value or both can be displayed at a time.
Chapter 4 Operation Control ■ Switching between Measured value, Trend Graph, and Statistical Processing Display 1 Press the [Up or Down] key and change the single measured value display to both value display screen. 2 Press the [Up or Down] key and select the OUT1 (upper part) or OUT2 (lower part) or Both values. 3 The selected item changes as shown below whenever the [SCREEN] key is pressed. OUT1 12.34560 UP 1.00000 STD 12.00000 LO –1.00000 NORMAL 512 OUT2 23.45670 OUT1 measured value UP 1.
Chapter 4 Operation Control ■ Trend Graph Display Trend graph can be displayed when both outputs are displayed. Trend Graph Display OUT1 23.45670 24.10000 22.90000 1M OUT2 10.00015 11.10000 4 8.90000 1M CAM1 RUN P – 00 AREA1 HEAD1 DIA AREA2 HEAD1 DIA A1 A2 2 2 01/02/18 17:50:42 ❈ The Y-axis of trend graph is automatically set for the next set value and displayed according to the tolerance set value.
Chapter 4 Operation Control Note: Graph data will be cleared in the following cases. • The LS-7600 Series is turned off. • The program number is changed, copied, or initialized. • Graph TIME data is changed. • The unit or minimum display unit is changed. The measured value will be displayed gray when the LS-7600 Series stands by. If the measured value exceeds the range of the Y-axis of the graph, the value will be displayed as the maximum or minimum value in the Y-axis direction.
Chapter 4 Operation Control 2 Press the [SCREEN] key for 2 s to return to the measured value display screen. NG RECORD Display NG RECORD NO TIME 1 01/01/01 12:00:00 2 01/01/01 11:55:00 3 01/01/01 11:43:00 4 01/01/01 11:39:00 5 01/01/01 11:15:00 6 01/01/01 10:53:00 7 01/01/01 10:47:00 8 01/01/01 10:32:00 9 01/01/01 10:32:00 10 01/01/01 09:37:00 4 Tips Note: P – 00 OUT1 P – 00 OUT1 P – 00 OUT2 P – 00 OUT2 P – 00 OUT2 P – 00 OUT2 P – 00 OUT1 P – 00 OUT1 P – 00 OUT2 P – 00 OUT2 RESULT 8.61150 8.60895 1.
Chapter 4 Operation Control Target Viewer Function A visual check on the image of each target is possible at a glance through the built-in image sensor camera of the measuring head. Clearance (AREA1) Target Outer diameter (AREA2) Measuring point 4 Center of measuring area Displays the center of the measuring area. OUT1 AREA1 Cursor Displays the set measurement range in AREA1. CAM1 RUN P – 00 12.34560 AREA2 Cursor Displays the set measurement range in AREA2. UPPER 1.00000 STANDARD 12.
Chapter 4 Operation Control Switching Cameras The CAMERA1 display on the upper right-hand side of the LCD monitor indicates the image sensor camera of measuring head 1. The CAMERA2 display indicates the image sensor camera of measuring head 2. ■ Single Value Display CAM1 CAM2 RUN P – 00 OUT1 Press the [SCREEN] key and select CAMERA1, CAMERA2, or both CAMERA1 and CAMERA2. 12.34560 UPPER 1.00000 STANDARD 12.00000 LOWER –1.
Chapter 4 Operation Control Hold Function Measured values can be held and displayed. Furthermore, MEA-Mode combinations allow the measurement of a variety of values, such as maximum and minimum values. The types of measured values held vary with the MEA-Mode. Refer to MEA-Mode on page 5-33 for details. Measured values are held and displayed through the following functional components.
Chapter 4 Operation Control Automatic Zero Function The automatic zero function sets measured values to zero (0.00000) instantly. This function is convenient for zero-point calibration for a variety of targets. Furthermore, master calibration is possible by using this function in combination with the offset function. Note: Tips This function will not be available while the controller is awaiting the result of measurement, during which the LCD monitor will display “– – –. – – – –.
Chapter 4 Operation Control Key Lock Function This function locks the control panel and remote control console, thus disabling key input. As a result, there will be no risk of making setting changes with careless key input. Key Lock Types The following types of panel locks are possible with the KEY LOCK settings as environment settings. Refer to Key Lock on page 6-13 for details. Set the key lock type to ALL: All keys excluding any key for the display change function will be locked.
Chapter 4 Operation Control Program Changes The following section provides information on how to call 16 programs saved in the LS7600 Series. Names and Number of Programs The selected program is displayed on the LCD monitor next to the [PROGRAM No.] key. Programs 0 through 15 (i.e., 16 programs in total) can be registered. CAM1 RUN P – 00 UT1 Display position 3.45670 PPER 1.00000 TANDARD 10.00000 OWER –1.00000 4 01/02/18 17 50 42 Programs are selected through the following functional components.
Chapter 5 Function Settings The LS-7600 Series is available to the measurement of a variety of items by making setting changes. These settings can be registered as programs that can be used whenever required. This section provides information on the overview of program settings and setting methods. Flow of Program Settings ............................................ 5-2 Default Values and Possible Setting Ranges .............. 5-4 Copying Program Setting Details and Initialization .....
Chapter 5 Function Settings Flow of Program Settings Area Settings ➾See page 5-9 Calibration Settings ➾See page 5-22 Calibration Settings (Optional) Measuring head Selection ■ Measuring head 1 (HEAD1) Measuring head 2 (HEAD2) 5 O Calibration mode selection Logic calibration Two-point calibration One-point calibration Area Selection DIA T-EDGE B-EDGE SEG Select the number of edges and mode (E/P) Out ■ Calibration Value Initialization M T O Optional Settings Focus monitor Area check ■ Differen
Chapter 5 Function Settings External I/O e page 5-22 ptional) Output Settings ➾See page 5-29 Optional Settings lection ■ tion libration libration ■ Area calculation Averaging number ■ Offset (in combination with the automatic zero function) ■ Abnormal value elimination ZERO alization TIMING Measuring Mode Settings RESET Tolerance Settings ➾See page 5-49 5 Output Settings ■ Analog output settings Scaling (Reference value/Threshold) Analog update mode Output waiting voltage Analog output ■
Chapter 5 Function Settings Default Values and Possible Setting Ranges The following programs settings are made by default.
Chapter 5 Function Settings ■ Output Settings (Page 5-29) Item Area calculation (CALC) Default OUT1: +1A1+0A2 OUT2: +0A1+1A2 Averaging number (AVE) 512 Measuring mode (MEA-MODE) Normal Offset 0.00000 Threshold Scaling at 10 V (A-OUT+10) Scaling at –10 V (A-OUT-10) Reference Scaling (A-SCALE) value Analog output setting: Analog update mode (A-SET) +30.
Chapter 5 Function Settings Copying Program Setting Details and Initialization The LS-7600 Series can be initialized so that the details of programs will return to ones set by default. Furthermore, the LS-7600 Series has a function making it possible to copy the details of any program to different programs. ■ Copying Program Setting Details The details of any program can be copied to different programs by using the P-COPY item as an environment setting. Refer to Copying Programs on page 6-5 for details.
Chapter 5 Function Settings Setting Screen The following section provides information on setting items. ■ Switching to Program Mode 1 Set the [Slide Switch] key to PROG mode. 2 By repeatedly pressing the [Up or Down] key, the cursor moves between LIMIT, AREA, OUTPUT, CALIB, and OPTION. 3 Locate the cursor to the desired item and press the [ENT] key to enter the mode. Tolerance Setting Mode OUT1 12.34560 UPPER STANDARD LOWER 1.00000 12.00000 —1.00000 OUT2 23.
Chapter 5 Function Settings Output Setting Mode CAM1 PROG CALC +1A1 +0A2 P – 00 LIMIT AVERAGE 512 AREA MEA-MODE NORMAL OFFSET 0.00000 OUTPUT CALIB OUT2 OPTION 23.45670 A1 2 CALC +0A1 +1A2 A2 2 AVERAGE 512 MEA-MODE NORMAL 01/02/18 OFFSET 0.00000 17:50:42 OUT1 12.34560 Calibration Setting Mode AREA1 THEORY 12.34560 T1-A T1-B T2-A T2-B 5.003000 5.000000 10.006000 10.000000 SET AREA2 THEORY 23.45670 T1-A T1-B T2-A T2-B 5 Option Setting Mode 5-8 5.003000 5.000000 10.006000 10.
Chapter 5 Function Settings Area Settings Area settings are used to determine the measurements taken from a variety of objects placed in the beam. The LS-7600 Series can use up to two measuring heads to set two measurements (AREA1 and AREA2) for simultaneous measurements. Area Settings Area settings include the following items. • Area: Select the measuring head number together with the DIA, T-EDGE, B-EDGE, or SEG mode. • Area display: Turn the cursor on the target viewer ON or OFF.
Chapter 5 Function Settings 4 Press the [Up or Down] key and move the cursor frame to the area to be set. To set 2, press the [Up or Down] key and move the cursor frame to AREA2. 5 Press the [ENT] key. Area Selection Area When a target is located in the measuring area as shown below, the light into the receiver is partly obstructed, thus creating a shadow. The contour of the shadow is called “edge.” The area is surrounded by the edge line.
Chapter 5 Function Settings ■ T-EDGE (Top Edge)/B-EDGE (Bottom Edge) Select either of these settings for gap measurement or positioning of rollers. When the LS-7600 Series is in T-EDGE measurement operation, it measures the width of the bright portion between the top of the measuring head and the first shadow in the measuring area.
Chapter 5 Function Settings E (Edge) Mode The LS-7600 Series measures objects based on the edges in this mode. Set the edge numbers to positive values beginning with the top of the measuring head. The following example shows positive edge numbers and measuring portions.
Chapter 5 Function Settings P (Pitch) Mode The LS-7600 Series measures objects based on the middle point between the edges of the objects in this mode. Top 1E 2E Target center pitch 3E 4E 5E 6E 1P 2P In this example as shown, five P-mode edges exist, based on the light and dark regions in the measuring area. Target center pitch: The space between 2P and 4P 3P 4P 5P 5 Bottom Edge 0 Edge 0 refers to a method of measurement, in which the center of the optical axis is used as an origin.
Chapter 5 Function Settings ■ Setting Procedure 1 Select HEAD and press the [ENT] key. The cursor moves to the setting item. 2 Press the [Up or Down] key and select the measuring head number. 3 Press the [ENT] key. The cursor moves to HEAD. 4 Select AREA and press the [ENT] key. The cursor moves to the setting item. 5 Press the [Up or Down] key and select 5 the measuring method. The following items are displayed for setting.
Chapter 5 Function Settings SEG (Segment) Setting (1) Press the [Up or Down] key and input the start edge (or start pitch) number. AREA1 HEAD AREA A-DISP FOCUS A-TEST (2) Press the [Left or Right] key and move the cursor to the end edge (or end pitch) number. AREA1 HEAD AREA A-DISP FOCUS A-TEST (3) Press the [Up or Down] key and input the edge pitch (or end pitch) number. AREA1 HEAD AREA A-DISP FOCUS A-TEST (4) Press the [ENT] key. The input values are entered and the cursor moves to AREA.
Chapter 5 Function Settings 2 Press the [Up or Down] key and select AREA1 HEAD AREA A-DISP FOCUS A-TEST ON or OFF. 3 Press the [ENT] key. The input item is entered and the cursor moves to A-DISP. 1 DIA ON 5 OFF Focus Monitor The focus monitor is a function that makes it possible to check whether the target is located within the range where the highest precision measurement of the target is ensured. The focus monitor range is the distance between the receiver and target in the measuring area.
Chapter 5 Function Settings Area Check Set the total number of edges to be measured. If the number of edges detected does not coincide with the set value, the measured value will be invalidated. This function makes it possible to prevent the measurement of abnormal values that may result if dust or oil is stuck on the measuring head. The LCD monitor will display the following data if the measured value is invalidated. OUT1 UPPER 1.00000 STANDARD 12.00000 LOWER –1.
Chapter 5 Function Settings Difference Check If there is a radical change between measured values, the change will be detected and output.. Setting Range • Difference average: 1 to 16 times • Difference value: 1 to 9999 μm Note: The results of difference checks will be output externally, provided that DIFF function output settings are made properly in advance. Refer to Function Output Selection on page 5-19 for details. The result output of a difference check is a 26.
Chapter 5 Function Settings Function Output Selection Select either Focus, Area, or Difference to be output from the output terminal block on the rear panel. Setting Range Any one of the following items is selectable for output. Refer to Section 7 I/O Terminals on page 7-1 for the FUNCTION output terminals in details. • Focus monitor • Area check • Difference check ■ Setting Method 1 Select FUNCTION and press the [ENT] key. The cursor moves to the setting item.
Chapter 5 Function Settings Change of Edge Detection Threshold The following section provides information on how to make threshold changes for the measurement of highly transparent objects.
Chapter 5 Function Settings ■ Setting method 1 Select LEVEL and press the [ENT] key. The cursor moves to the setting item. 2 Press the [Up or Down] key and input the set value. If the [Up or Down] key is pressed continuously, the set value will automatically increase or decrease. 3 Press the [ENT] key. AREA1 DIFF AVE LIMIT FUNCTION LEVEL 1 1000um FOCUS 50% AREA1 DIFF AVE LIMIT FUNCTION LEVEL 1 1000um FOCUS 50% The input value is entered and the cursor moves to LEVEL.
Chapter 5 Function Settings Calibration Settings The measured value of an target may have a delicate error due to the surface condition and angle of the target. Calibration settings correct for this error. It is possible to make calibration settings for each area independently. ■ Types of Calibration Settings Select the type of calibration from the following ones in advance. Refer to Calibration Type Selection on page 5-24.
Chapter 5 Function Settings ■ Procedure for Calibration Settings Take the following procedure if the remote control console is used. 1 Set the [Slide Switch] key to PROG. The LCD monitor changes to the PROGRAM SETUP screen from the RUN screen. 2 Press the [Down] key and move the cursor to CALIB. 3 Press the [ENT] key. A frame is displayed on the AREA1 side.
Chapter 5 Function Settings Calibration Type Selection The following section provides information on how to set the calibration type for each area. The procedure below is for AREA1 for a typical example. 1 With the AREA 1 box selected, press ENT. T1-A will be lit. Press the [Up] key and move the cursor to underline the type of calibration. “THEORY” The item is underlined and displayed. 2 Press the [ENT] key. The cursor appears on “THEORY”.
Chapter 5 Function Settings 1 While the cursor is at T1-A, press the [ENT] key and input the T1-A value. The display “*********” changes to “+00.00000.” The cursor moves to the “+” position and the value can be input. (1) Press the [Up or Down] key and select + or – as desired. (2) Press the [Left or Right] key and move the cursor between digits to input the value. AREA1 (3) Press the [Up or Down] key to input the value. (4) Repeat steps (2) and (3) until all the digits are input.
Chapter 5 Function Settings Two-point Calibration Settings Make two-point calibration settings as explained below. Setting Range Two-point calibration requires two types of master workpieces that are different to each other in size. If the master workpieces are target 1 (T1) and target 2 (T2) respectively, make settings so that the following conditions will be satisfied. T1-A < T2-A Furthermore, set the calibration span between 0.5 and 2.0.
Chapter 5 Function Settings 6 Press the [ENT] key. The input value is entered and the screen returns to the AREA selection status of calibration settings. Note: Tips If the calibration conditions are not satisfied, ERR1 will be displayed. Press the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the calibration settings again. When repeating the two-point calibration, the previous settings will appear on the screen, which will be different from those in the above setting example.
Chapter 5 Function Settings 5 Press the [ENT] key. The input value is entered and the screen returns to the AREA selection status of calibration settings. Note: Tips If the calibration conditions are not satisfied, ERR1 will be displayed. Press the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the calibration settings again. When repeating the one-point calibration, the previous settings will appear on the screen, which will be different from those in the above setting example.
Chapter 5 Function Settings Output Settings Values measured in AREA1 and AREA2 can be processed in a variety of ways and output as OUT1 and OUT2 signals. This section provides information on the details and methods of output settings. Possible Output Settings The following output settings are possible. • Area calculation: The measured values of two-area points specified by area settings are used for calculation processing. • Number of averaging measurements: Set the number of measuring times.
Chapter 5 Function Settings Making Output Settings for OUT1 (1) Press the [ENT] key. Making Output Settings for OUT2 (1) Press the [Down] key to move the cursor frame to OUT2. OUT1 CALC AVE MEA-MODE OFFSET (2) Press the [ENT] key. OUT2 CALC AVE MEA-MODE OFFSET CAM1 PROG +1A1 +0A2 P – 00 LIMIT 512 AREA NORMAL 0.00000 OUTPUT CALIB OPTION A1 2 +0A1 +1A2 A2 2 512 NORMAL 01/02/18 0.00000 17:50:42 Area Calculation The measured values from the two memory areas are used for calculation processing.
Chapter 5 Function Settings Obtaining the object width from Two Measuring Heads Set AREA1 for B-EDGE in measuring head 1 and AREA2 for B-EDGE in measuring head 2. A1 –1 A2 –1 ■ Setting Method 1 Press the [Up or Down] key and move the cursor to CALC. 2 Press the [ENT] key. The cursor moves to the coefficient for AREA1 (A1). 3 Press the [Up or Down] key and set the coefficient for AREA1 (A1). OUT1 4 Press the [Right] key and move the cursor to the coefficient for AREA2 (A2).
Chapter 5 Function Settings ■ Averaging Time and Refresh Cycle The LS-7600 Series averages measured values in a moving averaging method, thus making it possible to output the measured values within a short time. The following table lists the averaging time, the number of moving averaging data items, and the refresh cycle of a target according to the number of averaging measurements. 5 Number of averaging measurements Averaging time (ms) Number of moving data samples Refresh cycle (ms) 1 0.42 1 0.
Chapter 5 Function Settings Measuring Mode The Measuring mode allows the LS-7600 Series to capture and hold measured values. Setting Range The LS-7600 Series has 11 types of Measuring modes. Refer to the following information for the outline of each mode. Normal The measured value is displayed and output as it occurs. Bottom Hold The minimum value during the period specified is displayed and output. Average Hold The average value during the period specified is displayed and output.
Chapter 5 Function Settings Application Example of Measuring modes Select the measurement mode according to the type of target and measuring condition. The following applications are typical examples of measurement in a variety of Measuring modes. Application 5 Applicable RUN mode Continuous measurement of wire, tube, or fiber. Normal Measurement of maximum value within specified period. Peak Hold Auto Peak Hold Measurement of minimum value within specified period.
Chapter 5 Function Settings Timing Chart The following table lists timing charts for respective measuring modes.
Chapter 5 Function Settings ■ Peak Hold • Peak hold: The maximum value during the sampling period specified by external timing input is displayed and output as shown below. • Bottom hold: The minimum value during the sampling period specified is displayed and output. • Peak-to-peak hold: The maximum-to-minimum value during the period specified is displayed and output. • Average hold: The average value during the sampling period specified is displayed and output.
Chapter 5 Function Settings ■ Auto Peak Hold • Auto peak hold: The maximum value measured by the LS-7600 Series in continuous measurement operation is displayed and output. • Auto bottom hold: The minimum value measured by the LS-7600 Series in continuous measurement operation is displayed and output. • Auto peak-to-peak hold: The maximum-to-minimum value measured by the LS-7600 Series in continuous measurement operation is displayed and output .
Chapter 5 Function Settings ■ Sample Hold 1 The instantaneous value at the time specified by external timing input is displayed, and the value is output as shown below.
Chapter 5 Function Settings ■ Sample Hold 2 The value measured during the period specified by external timing input is displayed, and the value is output as shown below.
Chapter 5 Function Settings ■ Self-timing Hold The LS-7600 Series in self-timing hold operation will detect targets automatically with no external timing input. Then the measured value is displayed and output.
Chapter 5 Function Settings ● Self-timing Hold and Area Settings The LS-7600 Series in self-timing hold operation will detect a target when the target is within the optical axis. Then the LS-7600 Series will start sampling after a certain time period (i.e., a self-timing period) elapses and a certain number of edges are present based on the measurement mode. Therefore, the target may enter the measuring area from any direction as shown below.
Chapter 5 Function Settings Setting Method 1 Press [ENT] key (Down key) and move the cursor to MEA-MODE. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and OUT1 select the MEA-MODE from the following list. CALC +1A1 +0A2 AVE 512 MEA-MODE NORMAL OFFSET 0.00000 The following table lists available setting items with corresponding displays on the LCD monitor.
Chapter 5 Function Settings Offset An offset value is a desired value that can be added to or subtracted from measured values. If the offset value is set, it is possible to display the offset value immediately on the screen by using the auto zero function. Setting Range The range is initially set to –99.99999 to 99.99999mm. This range varies with the display unit or minimum display unit setting. Refer to Display Unit on page 5-57 and Minimum Display Unit on page 5-58.
Chapter 5 Function Settings Analog Output Settings The LS-7600 Series has analog voltage output according to the measured value. The range of analog voltage output is between –10 and +10 V. Setting Range The following analog output settings are available. • Scaling (reference value): In reference value mode (LIMITS), the deviation is output as a positive or negative voltage based on the reference value of 0 V.
Chapter 5 Function Settings The following table shows the relationship of scaling values along with their output ranges and resolutions. Scaling value [μm/V] Output range 1 ±10μm Resolution (❈) 0.01μm 2 ±20μm 0.01μm 5 ±50μm 0.02μm 10 ±100μm 0.04μm 20 ±200μm 0.08μm 50 ±500μm 0.2μm 100 ±1mm 0.4μm 200 ±2mm 0.
Chapter 5 Function Settings Setting Method 1 Press the [Up or Down] key and move the cursor to A–OUT+10 or A–OUT–10. 2 Press the [ENT] key. The cursor moves to the “+” position and the value can be input. 3 Press the [Up or Down] key and OUT1 select + or –. 4 Press the [Left or Right] key and move the carsor between digits to input the value. 5 Press the [Up or Down] key to input A-OUT+10 +30.00000 A-OUT -10 0.00000 A-SET OFF A-CONT +10.5V OUT1 the value.
Chapter 5 Function Settings ■ Output Waiting Voltage Take the following steps to set the voltage to be output when the display shows “---.-----.” 1 Press the [Up or Down] key and move the cursor to A-CONT. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and OUT1 select the setting item. 4 Press the [ENT] key. The input item is entered and the cursor returns to A-CONT. A-UNIT 1000um/V A-SET OFF A-CONT +10.
Chapter 5 Function Settings ■ Output Logic The logic of the comparator output and the WAITING output can be set to NO (normally open) or NC (normally closed). 1 Press the [Up or Down] key and move the cursor to OUT-SET. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and set the output logic. The following table lists available setting items with corresponding displays on the LCD monitor. LCD monitor OUT1 STOBE-T 0.4ms OUT-SET N.O. OFF-DLY OFF Setting N.O.
Chapter 5 Function Settings Tolerance (LIMIT) Settings The LS-7600 Series judges whether or not the measured values pass the criteria. The range of values used for judgment is called a tolerance.Tolerances can be set to OUT1 and OUT2 respectively. Types of Limits The following tolerance settings are possible. • Reference value Set the reference value and its upper and lower limits as deviation values. Measured values can be compared to three levels (i.e., HI, GO, and LO).
Chapter 5 Function Settings ■ Setting OUT1 Output (1) Press the [ENT] key. ■ Setting OUT2 Output (1) Press the [Down] key and move the cursor frame to OUT2. (2) Press the [ENT] key. OUT1 UPPER 1.00000 STANDARD 10.00000 LOWER —1.00000 OUT2 UPPER 1.00000 STANDARD 10.00000 LOWER —1.00000 CAM1 PROG P — 00 LIMIT AREA OUTPUT CALIB OPTION A1 2 A2 2 01/02/18 17:47:30 Limit Range The following ranges are used for the limits.
Chapter 5 Function Settings Limit Settings ■ Reference Value 1 Press the [ENT] key. The cursor is displayed at UPPER (upper limit). 2 Press the [ENT] key. The cursor moves to +, where numeric input is possible. 3 Press the [Up or Down] key and OUT1 select + or –. 4 Press the [Left or Right] key and move the cursor to the digit where the value is input. 5 Press the [Up or Down] key and UPPER +01.00000 STANDARD 10.00000 LOWER —1.00000 OUT1 select the value.
Chapter 5 Function Settings 3 Press the [Up or Down] key and OUT1 select + or –. 4 Press the [Left or Right] key and move the cursor to the digit where the value is input. 5 Press the [Up or Down] key and HH HI LO LL +40.00000 30.00000 0.00000 —10.00000 OUT1 select the value. 6 Repeat steps 4 and 5 to input all the digits. 7 Press the [ENT] key. HH HI LO LL +40.00000 30.00000 0.00000 —10.00000 The input items are entered and the cursor returns to HH.
Chapter 5 Function Settings ■ Setting Upper Limit Hold Value Take the following steps to set the upper limit hold value so that the LS-7600 Series will automatically detect and eliminate measured values exceeding the upper limit hold value. This function is enabled with the set value turned ON and disabled with the set value turned OFF. ● Turning OFF 1 Press the [Up or Down] key and move the cursor to HOLD-H. 2 Press the [ENT] key. The cursor moves to the setting item.
Chapter 5 Function Settings 08 09 10 Note: Press the [Up or Down] key and select the value. OUT1 Repeat steps (7) and (8) until all the digits are input. HOLD—H HOLD—L ON +90.00000 OFF Press the [ENT] key. The input item is entered and the cursor returns to HOLD-H. If there is any input value error, the value will not be entered but the cursor will return to the setting item.
Chapter 5 Function Settings Option Settings Settings, such as display unit and I/O terminal mode, are available on the LS-7600 Series. Option Settings The following option settings are available. • Tolerance mode: Sets the tolerance mode to use either reference values or thresholds. • Timing mode: Sets the timing input terminals for OUT1 and OUT2 respectively. • Unit: Changes the display unit of measured values. • Minimum display unit: Changes the minimum display unit of measured values.
Chapter 5 Function Settings Tolerance Mode (LIMITS) Take the following steps to select the appropriate tolerance mode. Setting Range The following modes are selectable. • Reference value mode: Use this mode if there is a workpiece size change with the deviations (tolerances) unchanged. • Threshold mode: Specify the upper and lower limits in absolute values. 1 Press the [Up or Down] key and move the cursor to LIMITS. 2 Press the [ENT] key. The cursor moves to the setting item.
Chapter 5 Function Settings 1 Press the [Up or Down] key and move the cursor to T-MODE. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and select the timing input. The following table lists available setting items with corresponding displays on the LCD monitor. LCD monitor Setting ASYNC Asynchronous mode SYNC Synchronous mode 4 Press the [ENT] key. The input item is entered and the cursor returns to T-MODE.
Chapter 5 Function Settings 1 Press the [Up or Down] key and move the cursor to D-UNIT. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and select the display unit. To change the unit to UNIT1 or UNIT2, press the SCREEN key for 2 s. 4 Press the [ENT] key. The input item is entered and the cursor returns to D-UNIT. LIMITS T-MODE D-UNIT D-RES C-OUT GRAPH-T OUT1 OUT2 STATS OUT1 OUT2 GROUP APP-MODE DEV ASYNC mm 0.
Chapter 5 Function Settings 3 Press the [Up or Down] key and select the minimum display unit. 4 Press the [ENT] key. The input detail is entered and the cursor returns to D-RES. LIMITS T-MODE D-UNIT D-RES C-OUT GRAPH-T OUT1 OUT2 STATS OUT1 OUT2 GROUP APP-MODE DEV ASYNC mm 0.00005 SUB 1m 1m OFF OFF OFF OFF I/O Mode The mode of the I/O connector on the rear panel of the controller can be changed. Setting Range Select the connector mode from the following ones.
Chapter 5 Function Settings Units, Default Minimum Values, and Setting Ranges Minimum display unit Setting item Calibration setting T1-A/B • T2-A/B Threshold Offset Output setting A– +10V A– –10V Threshold mode HH Option setting Ref. value Threshold Reference value mode Tolerance setting 5 0.00002mm 0.00005mm 0.0001mm 0.00001mm 0.02μm 0.05μm 0.1μm 0.
Chapter 5 Function Settings Minimum display unit Setting item Calibration setting T1-A/B • T2-A/B Threshold Offset Output setting A– +10V A– –10V Threshold mode HH Ref. value Threshold Reference value mode Tolerance setting Option setting 0.000001inch 0.00005inch 0.0001inch 0.0000005inch 0.02mil 0.05mil 0.1mil 0.
Chapter 5 Function Settings Graph TIME Take the steps below to set the display periods of trend graphs for OUT1 and OUT2 respectively. For details, refer to Trend Graph Display on page 4-4. Setting Range Select the periods from the following. 1m, 10m, 30m, 1H, 6H, 12H and 24H. 1 Press the [Up or Down] key and move the cursor to OUT1 of GRAPH-T. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and select the display period for OUT1 of GRAPH-T.
Chapter 5 Function Settings 3 Press the [Up or Down] key and select the statistical processing mode. The following table lists available settings items with corresponding displays on the LCD monitor. LCD monitor Setting OFF Statistical processing OFF EXT External sync 1000 Internal sync LIMITS T-MODE D-UNIT D-RES C-OUT GRAPH-T OUT1 OUT2 STATS OUT1 OUT2 GROUP APP-MODE DEV ASYNC mm 0.00005 SUB 1m 1m OFF OFF OFF OFF ■ Internal Sync Selected (1) Press the [ENT] key.
Chapter 5 Function Settings 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and select the set value. If the [ENT] key (Up or down key) is pressed continuously, the set value will automatically increase or decrease. 4 Press the [ENT] key. The input item is entered and the cursor moves to GROUP. LIMITS T-MODE D-UNIT D-RES C-OUT GRAPH-T OUT1 OUT2 STATS OUT1 OUT2 GROUP APP-MODE DEV ASYNC mm 0.
Chapter 5 Function Settings Group Processing The LS-7600 Series calculates average values from measured data on groups of workpieces. The average values are used to determine the measured values. Furthermore, group values can be statistically processed. Each workpiece within a group is measured by OUT1. The measured results of OUT1 are processed by OUT2. The following measurement modes can be used for group processing.
Chapter 5 Function Settings Statistical Processing The statistical processing function calculates peak, bottom, average, peak-to-peak, standard deviation, N output, HH output, HI output, GO output, LO output, and LL output values. Statistical Processing Items Available • Statistical processing is possible independently for each program number and OUT (output). • Statistical processing under RS-232C control is possible. • The results of statistical processing can be printed out.
Chapter 5 Function Settings 5 The result of statistical processing is displayed. Internal Synchronous Setting 1 Press the [STATS] key or turn ON the STATS input. 2 The previous statistical processing result is reset and “STATS” is displayed. 3 Measure the workpiece. 4 When the measurement result of the specified number of workpieces is obtained, the result will be displayed.
Chapter 5 Function Settings Application Function The application mode consists of the drill function and the dual head function. Drill Function The drill function can measure the external diameters of objects, such as drills or gears with odd-numbered teeth. The diameter of the circumscribed circle made by the turning object is measured. Circumscribed circle (External dia.) 5 ■ Operation Method 1 Set the APP-MODE to DRILL.
Chapter 5 Function Settings Note: • OUT2 cannot be used. • The following items will be automatically set when APP-MODE is set to DRILL. Set value Set item Area 1 Area 2 HEAD 1, 2 1, 2 AREA +2E~0E 0E~–2E Area setting OUT1 CALC A1: +1 A2: 0 Setting AVERAGE possible Output setting Peak hold or auto MEA-MODE peak hold OFFSET GROUP Option setting APP-MODE Setting possible OFF Remarks Select the measuring head number to be used. Automatically set to a fixed value.
Chapter 5 Function Settings Dual Head Function The dual head function uses two measuring heads to measure the width of sheet material or the outer diameter of thick targets that are too large to be measured with a single head. ■ Mounting the Measuring Heads When using the dual head function, mount the two measuring heads so that their top sides face each other. Top ■ Operation Method 1 Set the unit and the minimum display unit according to the size of the object.
Chapter 5 Function Settings Note: The following items will be automatically set when APP-MODE is set to DUAL. Set value Set item Area 1 Area setting HEAD 1 1 AREA +0E~–2E +0E~–2E OUT1 Output setting CALC Option setting APP-MODE A1 : +1 A2 : +1 Remarks Area 2 Automatically set to a fixed value. The set value cannot be changed. OUT2 A1 : +1 A2 : +1 Automatically set to a fixed value. The set value cannot be changed.
Chapter 5 Function Settings MEMO 5 5-72
Chapter 6 Environment Settings This section provides information on environment settings that include RS-232C interface settings along with basic operation settings and changes. Possible Environment Settings .................................... 6-2 Details of Environment Settings ..................................
Chapter 6 Environment Settings Possible Environment Settings The following environment settings are possible. • Program: Program setting changes, program copying, and setting initialization. • RS-232C: Baud rate, parity, stop bit, data length, data mode, and data transmission mode of the RS-232C interface. • Head: LED inspection function and mutual interference prevention function. • Beep: Turn the key buzzer ON or OFF. • Key lock: Select which keys are locked.
Chapter 6 Environment Settings ■ Head (See page 6-9) Item Setting range Default LED inspection function (L-INSPCT) (–) 1 or 2 (Connected head number) Mutual interference prevention function (H-SYNC) OFF OFF, ON ■ Buzzer (See page 6-12) Item Setting range Default Buzzer ON/OFF switch (BEEP) ON OFF, ON ■ Key Lock (See page 6-13) Item Setting range Default Key lock function (KEY-LOCK) ALL ALL, PART ■ Displayed Language (See page 6-13) Item Displayed language selection (LANGUAGE) 6 Settin
Chapter 6 Environment Settings Changing Environment Settings Make changes to the environment settings. 1 Set the [Slide Switch] key to PROG. 2 Press the [ESC] key for two seconds. The LCD monitor changes from the program setup screen to the environment setup screen.
Chapter 6 Environment Settings Details of Environment Settings This section provides information on environment setting items and setting methods. Program The LS-7600 Series ensures ease of program settings or changes. Possible Settings • Program Selection (P-SELECT): Select the program from the front panel or the rear terminals. • Copying Programs (P-COPY): Program can be copied to other programs. Furthermore, a selected program can be initialized.
Chapter 6 Environment Settings 1 Press the [Up or Down] key and select P-COPY. 2 Press the [ENT] key. The cursor moves to the number of the program to be copied. 3 Press the [Up or Down] key and select the number of the program to be copied or “EE”. 4 Press the [Left or Right] key and move the cursor to the item of the destination program number. 5 Press the [Up or Down] key and select the destination program number. 6 Press the [Left or Right] key and move the cursor to the copy confirmation item.
Chapter 6 Environment Settings RS-232C Take the following steps to make the communication settings for the RS-232C interface and relative settings for peripheral devices. Tips Communication settings for the RS-232C interface vary with the peripheral device to be connected or the type of output desired. For the details of RS232C, refer to Chapter 8 RS-232C on page 8-1. Setting Items The following items can be set.
Chapter 6 Environment Settings 3 Press the [Up or Down] key and set the parity. The following table lists available setting items with corresponding displays on the LCD monitor. Setting item Description NONE None EVEN Even ODD Odd 4 Press the [ENT] key. The input item is entered and the cursor returns to PARITY. ■ Stop Bit 1 Press the [Up or Down] key and select STOPBIT. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key and select the set value.
Chapter 6 Environment Settings 3 Press the [Up or Down] key and select the data mode. The following table lists available setting items with corresponding displays on the LCD monitor. Setting item Description NORMAL Only the item OFF is selectable in data transmission mode. PLC The following environment settings are fixed. Baud rate: 9600 Data length: 8 Stop bit length: 1 Parity: Even PRINTER1/PRINTER2 The items S1, S2, or S3 is selectable in data transmission mode. 4 Press the [ENT] key.
Chapter 6 Environment Settings Check on Intensity of LED Light Source With the LED inspection function, the distribution of light will appear on the LCD monitor as shown below, which makes it possible to check the change of light intensity after the LS-7600 Series is installed. 100% The distribution of light is uniform. 100% 100% 100% 100% Indicates that 100% light is distributed. (Varies by a few percent according to the unit.) 100% 100% 100% This bar indicates 5% intensity.
Chapter 6 Environment Settings Check on Fluctuation of Light Intensity Due to Dirt or Dust 100% The distribution of light has fluctuation. 100% 95% 95% 90% 90% 90% 95% In the above case, refer to page 3 and clean the cover glass. Note: LED Inspection Function • The function operates correctly, provided that the position of the measuring head remains the same before, while, and after the intensity of light is checked. The intensity of light fluctuates if the mounting position shifts.
Chapter 6 Environment Settings ■ Mutual Interference Prevention Function This function prevents mutual interference that may result when two measuring heads are located at right angles. With the function used, the LED light sources are turned sequenced ON and OFF. Note: If the mutual interference prevention function is used, the following settings will change. • The sampling cycle of measurement will be set to 1,200 times per second.
Chapter 6 Environment Settings Key Lock Take the following steps to specify the key lock range when the key lock function is set. Setting Range Set the following items. • All (ALL): All function keys other than the display selection function key are locked. • Part (PART): All function keys other than the display selection and tolerance setting function keys are locked. Refer to page 4-11 for the key lock function in detail. 1 Press the [Up or Down] key and select KEY LOCK. 2 Press the [ENT] key.
Chapter 6 Environment Settings LCD Monitor Backlight Take the following steps to turn the backlight of the LCD monitor ON and OFF. Note: The backlight will be OFF if there is no key input for 10 minutes. To turn ON the backlight again, press any control panel key or remote control console key. 1 Press the [Up or Down] key and select AUTO OFF. 2 Press the [ENT] key. The cursor moves to the setting item. 3 Press the [Up or Down] key to turn the function ON or OFF.
Chapter 7 I/O Terminals This section provides information on I/O terminal specifications and timing charts. Nomenclature and Functions of I/O Terminals ............ 7-2 Timing Chart .............................................................
Chapter 7 I/O Terminals Nomenclature and Functions of I/O Terminals This section provides nomenclature and functions of the I/O terminals of the LS-7600 Series. Types of I/O Terminals The LS-7600 Series has three types of I/O terminals. Use them according to the function used and the environment.
Chapter 7 I/O Terminals I/O Signal Names and Descriptions The I/O terminals handle the following signals. Refer to Function on page 7-7 for the functions of the signals in detail. For the electrical specifications of signals, refer to the reference pages indicated in the table.
Chapter 7 I/O Terminals Connector I/O Connector I/O signals operate in either SUB mode or BCD. Refer to I/O Mode on page 559 for details. Applicable Connector Specifications HIROSE ELECTRIC’s HIF3B Series or equivalent ones Terminal number 1 21 2 22 3 23 4 24 5 25 6 26 7 27 8 28 9 29 10 30 11 31 12 32 13 33 14 34 15 35 7 16 36 17 37 18 38 19 39 20 40 ■ I/O Signal Names and Descriptions The connector I/O terminals handle the following signals.
Chapter 7 I/O Terminals SUB Mode Terminal number 1 21 2 22 3 23 4 24 5 25 6 26 7 27 8 28 9 29 10 30 11 31 12 32 13 33 14 34 15 35 16 36 17 37 18 38 19 39 20 40 Signal name COM2 (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) COM2 (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) (Not used) OUT1 STATS RUN OUT2 STATS RUN (Not used) AREA2 FUNCTION OUT2 HH OUT2 HI OUT2 GO OUT2 LO OUT2 LL OUT2 WAITING
Chapter 7 I/O Terminals BCD Mode Terminal number 1 21 2 22 3 23 4 24 5 25 6 26 7 27 8 28 9 29 10 30 11 31 12 32 13 33 14 7 34 15 35 16 36 17 37 18 38 19 39 20 40 Signal name COM2 BCD POLE BCD DIGIT 7 (8) BCD DIGIT 7 (4) BCD DIGIT 7 (2) BCD DIGIT 7 (1) BCD DIGIT 6 (8) BCD DIGIT 6 (4) BCD DIGIT 6 (2) BCD DIGIT 6 (1) BCD DIGIT 5 (8) BCD DIGIT 5 (4) BCD DIGIT 5 (2) BCD DIGIT 5 (1) COM2 BCD DIGIT 4 (8) BCD DIGIT 4 (4) BCD DIGIT 4 (2) BCD DIGIT 4 (1) BCD DIGIT 3 (8) BCD DIGIT 3 (4) BCD DIGIT 3 (2) BCD DIGIT 3
Chapter 7 I/O Terminals VIDEO Output The LS-7600 Series can output the video signal of the LCD monitor image on the controller, and display it on a large screen. Applicable Connector: RCA connector (plug) Functions The following section provides information on I/O signal functions. For the arrangement of each terminal, refer to pages 7-2 and 7-4. ■ Alarm Output Name ALARM Description Reference page Turns ON when an error is detected.
Chapter 7 I/O Terminals ■ Program Number Selection Input Name P1 through P4 Input Description Reference page Externally changes program numbers. 7-3 The following table shows a list of program numbers and the status of the P1 terminal through the P4 terminal.
Chapter 7 I/O Terminals ■ BCD Output Name Description Reference page BCD POLE output BCD polarity output. OFF: Positive ON: Negative 7-6 BCD DIGIT output BCD numeric output for each digit. Negative logic (i.e., Turns ON (set to 1) when NPN open connector output is ON). Comparator waiting status (with "–" displayed for each digit): All digits set to B (1011) Overrange (with "F" displayed for each digit): All digits turn ON (1111). 7-6 BCD INDICATE OUT1/2 output Present OUT number turns ON.
Chapter 7 I/O Terminals ■ NPN Open Collector Output 1 Internal circuit +5V +5V 470Ω Output 2.2kΩ 10kΩ 10nF COM1 Max. applied voltage 40V Max. sink current 100mA Residual voltage 1 V max. Current leakage 0.1 mA max. Max. applied voltage 30V Max. sink current 30mA Residual voltage 1 V max. Current leakage 0.1 mA max. ■ NPN Open Collector Output 2 Internal circuit +5V +5V 470Ω Output 10kΩ 10kΩ 10nF COM2 7 Internal circuit ■ Analog Voltage Output 100Ω Output range ±10.
Chapter 7 I/O Terminals Timing Chart This section provides information on the timing charts of BCD output in a variety of measuring modes. ■ Normal, Auto Peak Hold, Auto Bottom Hold, and Auto Peak-to-peak Hold Refresh cycle ON Comparator output OFF ON WAITING output OFF ON BCD output OFF STROBE time ON STROBE output OFF TIMING input 0±0.1ms ON OFF Min 2.1ms Min2.1ms Min Min2.1ms Max2.5ms ON RESET/ZERO input OFF Max2.5ms 2.1ms Max2.5ms ON P1 input to P4 input OFF RESET Min Min6.0ms time 6.
Chapter 7 I/O Terminals ■ Sample Hold 2 ON OFF ON WAITING output OFF ON BCD output OFF Comparator output 0±0.1ms ON STROBE output OFF Min Min2.1ms 2.1ms ON TIMING input OFF RESET/ZERO input STROBE time Max2.5ms ON OFF Sampling Min2.1ms Min2.1ms period Max2.5ms ON P1 input to P4 input OFF Min6.0ms Min6.0ms Max12.0ms • Triggered by the acknowledgment of timing input, the measured value will be determined with an elapse of the sampling period (i.e., reset time).
Chapter 7 I/O Terminals ■ BCD Output Switching Refresh cycle OUT1 Measured value OUT1 STROBE output ON OFF Refresh cycle OUT2 Measured value OUT2 STROBE output BCD SELECT input BCD DATA output BCD INDICATE output ON OFF ON OFF 0±0.1ms ON OFF ON OUT1 OFF ON OFF OUT1 OUT1 OUT2 Max2.1ms OUT2 OUT1 OUT1 Max2.1ms • BCD output can be set to OUT1 or OUT2 according to BCD SELECT input.
Chapter 7 I/O Terminals ■ Minimum Input Period Refer to the relevant timing charts for the delay time (response delay time) in response to input. Name Specifications ON : 2.1ms TIMING/ZERO/RESET OFF : 2.1ms ON : 6.0ms P1~P4 OFF : 6.
Chapter 8 RS-232C This section provides information on how to connect the LS7600 Series to external devices along with RS-232C communications functions used for the external control of the LS-7600 Series. Specifications .............................................................. 8-2 Communication using Serial Commands .................... 8-4 Communication using an External Synchronous Trigger ....................................................................... 8-26 Command List .....................
Chapter 8 RS-232C Specifications The LS-7600 Series has RS-232C communications functions. The following section provides information on the RS-232C communications specifications of the LS-7600 Series and the outline of communications settings that vary with the connecting device.
Chapter 8 RS-232C ■ RS-232C Connection Settings Make the following environment settings according to the connecting device. For details, go to the reference pages indicated in the table. Refer to page 6-7 for RS-232C environment setting details. External device Non-protocol communications with PCs Environment setting item D-MODE D-SEND NORMAL or PLC link units.
Chapter 8 RS-232C Communication using Serial Commands To enable the LS-7600 Series to transmit measured values or allow setting changes, connect a peripheral device, such as a PC or PLC link unit to the LS-7600 Series. Connecting to a PC/PLC Link Unit The following section provides information on how to connect PCs or PLC link units to the LS-7600 Series. Use the dedicated OP-35382 cable (with 9-pin D-sub connectors) or the OP-25253 (with 25-pin D-sub connectors) for connection.
Chapter 8 RS-232C Connecting to KV Series The following information provides information on how to connect Keyence’s KV-series Programmable Logic Controller to the LS-7600 Series. Use the dedicated OP-25254 to connect the KV to the 7600 Series. Note: • The output of the LS-7600 Series under external synchronous control cannot be connected to printers. • Only the KV-300/Visual Series can transmit text data. The KV-10/80 series can not be used with this mode. Refer to page 8-26 for more details.
Chapter 8 RS-232C Outline of Command Formats When LS-7600 Series receives a command from the external device, the LS-7600 Series will respond to the receuved command and return a response to the external device. The following section provides information on respective command formats. Note: When you write a control program, check that the program will enable the external device to send the next command after the external device confirms the response from the LS-7600 Series.
Chapter 8 RS-232C Measurement Serial Commands The following section provides information on serial commands used for measurement control.
Chapter 8 RS-232C Commands in Detail The following section provides information on commands and responses that the LS-7600 Series returns when commands are correctly processed. For error responses, refer to Outline of Command Formats on page 8-6. Refer to Timing Chart on page 8-24 for response delays. ■ Measured Value Output The LS-7600 Series can output a measured value. The following types of measured value output are possible. • Measured value output: The measured value is output.
Chapter 8 RS-232C Response (in case of statistical processing data) M q [ , , , , , , , , , , , , , , , , , , , , , , s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s u u u u u u s s s s s s s s s s s Output number MAX MIN AVE PP Deviation Number of da
Chapter 8 RS-232C Peak Hold, Bottom Hold, Peak-to-peak Hold, Average Hold, or Sample Hold 1 Measuring mode Sampling period Sampling period Max Max Measured value Command Response U[] L[] U[] U[] Measured value L[] U[] Measured value In peak hold mode, the maximum value of the sampling period is returned as a measured value. When the LS-7600 Series receives command L (measured value output), the internal timing is triggered and the determined measured value will be returned.
Chapter 8 RS-232C ■ Timing ON This command turns timing input ON. Command H q cr With command H0 executed, the operation of the LS-7600 Series will be the same as that with SYNC set as a T-MODE environment setting item. LS Response H q cr Output number 0: OUT1+OUT2 1: OUT1 only 2: OUT2 only With command H1 or H2 executed, the operation of the LS-7600 Series will be the same as that with ASYNC set as a T-MODE environment setting item. ■ Timing OFF Command U q cr This command turns timing input OFF.
Chapter 8 RS-232C ■ Statistical Sampling ON This command starts data sampling for statistical processing. Command Output number 0 : OUT1+OUT2 1 : OUT1 2 : OUT2 O q cr LS Response O q cr ■ Statistical Sampling OFF This command completes data sampling for statistical processing. Command Output number 0 : OUT1+OUT2 1 : OUT1 2 : OUT2 R q cr External synchronization: The LS-7600 Series completes statistical sampling and starts statistical processing.
Chapter 8 RS-232C ■ Function Output This command makes it possible to monitor the status of the FOCUS output and the status of the area check output.
Chapter 8 RS-232C Setup Serial Commands The following section provides information on command formats that are used to make setting changes.
Chapter 8 RS-232C ■ Numeric Setting Format ( m m m m m m m m Format) This format is a fixed length format consisting of a single mark character and seven numeric characters with no decimal used. The following format will be used for the corresponding measured value. Measured value Format 1.23450 + 0 1 2 3 4 5 0 Commands in Detail The following section provides information on commands and responses that the LS7600 Series returns when commands are correctly processed.
Chapter 8 RS-232C ■ Area Check This command makes area check ON/OFF settings and changes the number of edges. Command S D , A T , a , f f f cr Area number 1 : AREA 1 2 : AREA 2 LS Number of area check edges OFF : OFF 002~127 : Value specified with the area check function turned ON. Response S D , A T , a cr Area number ■ Level This command changes threshold level settings.
Chapter 8 RS-232C ■ Measuring Mode This command changes measuring mode settings. Measuring mode 00 : NORMAL 04 : AV-HOLD 01 : P-HOLD 05 : A-P-HOLD 02 : B-HOLD 06 : A-B-HOLD 03 : D-HOLD 07 : A-D-HOLD Command S D LS , M E , h , k k cr Output number 1 : OUT1 2 : OUT2 Response S D , M E 08 : S-HOLD1 09 : S-HOLD2 10 : SELF-T , h cr ■ Self-timing Period If the measuring mode is set to self-timing hold, this command will change self-timing period settings.
Chapter 8 RS-232C Threshold Mode +10V set value Value setting format (See page 8-15) Command S D LS , A O , h , , A O , m m m m m m m m cr Output number 1 : OUT1 2 : OUT2 Response S D m m m m m m m m –10V set value Value setting format (See page 8-15) , h cr ■ Tolerance This command changes tolerance settings. The LS-7600 Series will return an error response if a command received is different in tolerance mode.
Chapter 8 RS-232C ● Threshold Mode HH Tolerance Value setting format (See page 8-15) Command S D LS , H H , h , Output number 1 : OUT1 2 : OUT2 Response S D , H H m m m m m m m m cr , h cr HI Tolerance Value setting format (See page 8-15) Command S D LS , H I , h , Output number 1 : OUT1 2 : OUT2 Response S D , H I m m m m m m m m cr , h cr LO Tolerance Value setting format (See page 8-15) Command S D LS , L O , h , Output number 1 : OUT1 2 : OUT2 Response S D , m m
Chapter 8 RS-232C ■ Hold Value Elimination ON/OFF This command turns the function of abnormal value elimination ON and OFF. Command S D , J O , h , Output number 1 : OUT1 2 : OUT2 LS Lower limit ON/OFF 1 : OFF 2 : ON w w cr Upper limit ON/OFF 1 : OFF 2 : ON Response S D , J O , h cr ■ Hold Value Elimination Set Value Change the set value of the function of abnormal value elimination.
Chapter 8 RS-232C ■ Group Processing This command turns the group processing function ON and OFF and makes set value changes. Command S D LS , G R , p p p cr Group processing setting OFF : OFF ON, Number of groups : 002~255 Response S D , G R cr ■ Statistical Processing This command turns the statistical processing function ON and OFF and changes the number of processing data items.
Chapter 8 RS-232C Serial Commands to check the LS Settings The following section provides information on command formats used to check setting details.
Chapter 8 RS-232C Batch Commands It is possible to read the whole setting data of the LS-7600 Series and save the data on the PC. The saved data on the PC can be written to the controller as well. Use the program read and program write commands. These commands allow the following operations. • Reading and writing program settings (auto zero set values are included). • Reading and writing environment settings.
Chapter 8 RS-232C ■ Reading Environment Settings It is possible to read all environment settings in detail. Note: P-COPY, P-CLEAR, or L-INSPCT settings are not read with this command. Command S A cr LS Response S A , Data 0 Data 1 Data X Checksum Checksum Binary data Checksum range Data 0 through data X (X=87) The checksum need not be calculated for simple upload and download operations.
Chapter 8 RS-232C ASCII Table Rightmost 4 bits Leftmost 4 bits 0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7 8 9 A B C D E F N D S D U L O L E C H 1 S D T C X 2 E D T C X 3 E D O C T 4 E N N A Q K A S C Y K N B E E T L B B CA S N H E T M L SU F B V ES T C F F C R S O S I S P 0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z a b c d e f g h i j k \ l m n o p q r s t u v w x y z D T 8 8-25
Chapter 8 RS-232C Communication using an External Synchronous Trigger The following section provides information on how to obtain output from the RS-232C interface in synchronization with a timing input into the input terminal. This operation does not require ASCII commands.
Chapter 8 RS-232C Timing Chart The following timing chart applies to timing input and RS-232C data output (the measuring mode is not set to sample hold 2 or self-timing hold.) TIMING input ON OFF RS-232C data output Measured value 10ms • If the measuring mode is set to normal, auto peak hold, auto bottom hold, or auto peak-to-peak hold, the measured value will be transmitted at the moment timing input is turned ON.
Chapter 8 RS-232C Simultaneous Measured Value The following command format will be used if SYNC is set as a T-MODE option setting.
Chapter 8 RS-232C Command List Measurement Serial Commands Command Measured value output (Page 8-8) M q , r cr r cr M R cr Measured value output under timing control (Page 8-9) Timing ON (Page 8-11) Timing OFF (Page 8-11) Reset (Page 8-11) Auto zero ON (Page 8-11) Auto zero OFF (Page 8-11) Statistics ON (Page 8-12) Statistics OFF (Page 8-12) Program number selection (Page 8-12) Program number check (Page 8-12) Key lock ON/OFF (Page 8-12) Function output (Page 8-13) L q , Response M q [ , , L q
Chapter 8 RS-232C Setup Serial Commands Output setting Area setting Command S D , H E , a , b cr S D , H E , a cr S D , A R , a , c S D , A R , a cr S D , S E , a , d d d d e S D , S E , a cr S D , A T , a , f S D , A T , a cr S D , L E , a , g g cr S D , L E , a cr Calculation (Page 8-16) Average (Page 8-16) RUN mode (Page 8-17) Self-timing (Page 8-17) Offset (Page 8-17) Analog Reference value (Page 8-17) Threshold (Page 8-18) S D , C A ,
Chapter 8 RS-232C Serial Commands to check the LS Settings Output setting Area setting Command S C , H E , a cr S C , H E , a , b cr S C , A R , a cr S C , A R , a , c S C , S E , a cr S C , S E , a , d d d d e S C , A T , a cr S C , A T , a , f S C , L E , a cr S C , L E , a , g g cr Calculation (Page 8-16) Average (Page 8-16) RUN mode (Page 8-17) Self-timing (Page 8-17) Offset (Page 8-17) Analog Reference value (Page 8-17) Threshold (Page 8-1
Chapter 8 RS-232C Batch Commands Command 8 8-32 Read Program Settings S R cr Writing Program Settings S W , Reading Environment Settings S A cr Writing Environment Settings S B , Data 0 Data 1 Data X Checksum Checksum Data X Checksum Checksum Data 0 Data 1 Data X Checksum Checksum S B cr 8-23 8-23 S W cr S A , Data 0 Data 1 Page Response S R , Data 0 Data 1 Data X Checksum Checksum 8-24 8-24
Chapter 9 Specifications This section provides the specifications, characteristics, and external dimensions of the LS-7600 Series. Specifications .............................................................. 9-2 Characteristics ............................................................ 9-4 Mode Specifications .................................................... 9-7 External Dimensions ...................................................
Chapter 9 Specifications Specifications Controller Model LS-7601 Number of connectable measuring heads 2 (interchangeable) Display Display type TFT5.5-type LCD Backlight: Cold cathode fluorescent lamp (with an average life of 40,000 hours) Minimum display unit 0.01µm to 100µm (7 selectable levels) Display range ±99.99999~±9999.
Chapter 9 Specifications Measuring Head Model LS-7010M LS-7030M LS-7070M Measuring range 0.04 to 6mm 0.3 to 30mm 0.5 to 65 mm Minimum detectable object 0.04mm 0.3mm 0.5 mm T-to-R distance 60±5mm 160±40mm 250±50 mm CCD scanning range Approx. 7mm GaN green LED Approx. 33 mm Approx. 69 mm Measurement position accuracy ±0.5µm (❈1) ±2µm (❈3) ±3µm (❈5) Repeating accuracy ±0.06µm (❈2) ±0.15µm (❈4) ±0.
Chapter 9 Specifications Characteristics Measurement Range and Accuracy LS-7010M 40 30±5 81 (30) Receiver Transmitter 4 6 ±0.5μm ±1.
Chapter 9 Specifications LS-7070M 150 56 100±50 Receiver Transmitter ±3μm 65 40 ±6μm 20 30 Unit : mm Temperature Characteristics The following table shows typical examples of the temperature characteristics of the LS-7600 Series. LS-7010M Temperature 0˚C (32˚F) Drift 0.25 10˚C (50˚F) 20˚C (68˚F) 30˚C (86˚F) 40˚C (104˚F) 50˚C (122˚F) 0.10 0.00 -0.05 -0.10 -0.
Chapter 9 Specifications Response Deley Time ■ Measurement Response Time The following table lists the response time between the moment the dimensional change of a target is detected and the measured result is obtained. Averaging times Measurement response time (ms) Measurement averaging time (ms) Refresh cycle (ms) Internal processing time (ms) 1 3.34 0.42 0.42 2.5 2 4.16 0.83 0.83 2.5 4 5.00 1.67 0.83 2.5 8 6.66 3.33 0.83 2.5 16 10.00 6.67 0.83 2.5 32 16.66 13.33 0.83 2.
Chapter 9 Specifications Mode Specifications ■ RUN Mode Status Measured value display RS-232C Output Measuring Numeric display Comparator display 5-level discrimination WAITING STROBE Analog output BCD ALARM FUNCTION Reception Response Comparator +All Comparator – – – – – – – OFF OFF ON OFF waiting O O digits: B waiting voltage Positiveside range over Value Negative side range over +FFFFFFF OFF Measured value Comparator output Comparator output OFF Reference value: LO Reference value: LO Threshol
Chapter 9 Specifications External Dimensions It is possible to download CAD data on the external dimensions of the LS-7600 Series from KEYENCE’s URL (http://www.keyence.com/). ■ LS-7601 Controller Panel Mounting Dimensions (191) (Outer size of mounting bracket) 175 22.3 5 60 111 (Effective display size) 207 Panel thickness 1.2~4.0 169 Mounting screw 23.2 25 83 (Effective display size) 60 Mounting bracket 124 130 +1 0 +1 0 200 (65 ) ■ LS-7010M Measuring Head 4-ø4.
Chapter 9 Specifications ■ LS-7030M Measuring Head 4-M2.5 10 Depth: 3 max. 24 24 4-ø4.5 through screw holes ø8 depth 4.5 15 45 22±0.2 Measuring position 34 109 55 6 55 Measuring range 71.5 60 126.5 30 4-M2.5 Depth: 3 max. 45 26 38 13.5 Front view of receiver 10 4-M4 Depth: 10 max. 15 13.5 Front view of transmitter 60 71.5 35.5 5.5 5.5 3-ø5.
Chapter 9 Specifications ■ Remote Control Console R2 44 23 3 12 18 12 103 6.5 ø11 ø4.5 Cable length: 3m ■ LS-S11 AC Power Supply Stand 171 196 60 (35) 14 178 43.5 148 216 (100) 201 ■ Extension Cable 9 LS-C**A Extension Cable (Cable between the controller and measuring head) 43 ø6.3 43 Unit: mm 14.7 14 LS-C**AM Camera Cable 35 ø4.8 35.3 11.5 11 OP-42182/OP-42183 TR Cables 35 11.5 9-10 ø4.8 35.
Chapter 10 Troubleshooting This section provides information on troubleshooting and the contents of error messages. Troubleshooting Guide .............................................. 10-2 Error Messages .........................................................
Chapter 10 Troubleshooting Troubleshooting Guide If the LS-7600 Series fails to operate, follow the inspection steps provided below. If the LS7600 Series does not operate normally after the inspection, contact your KEYENCE representative. Problem Inspection Remedy If 24 VDC is used Check that the power cable is Connect the power cable correctly. connected correctly. (See page 1-18) Check that the power supply is Use a power supply within the specified within the specified range. range.
Chapter 10 Troubleshooting Problem The measured value is not displayed. Inspection Remedy Check that the LED light source of Check the connection of the cable the transmitter is lit. between the transmitter and receiver. (See page 1-18) If the LED light source is still OFF, the LED light source needs a replacement. Contact your KEYENCE representative. Check that the measuring head is Connect the cable correctly. connected correctly.
Chapter 10 Troubleshooting Problem Inspection Remedy Check that the cover glass of the transmitter Remove the dirt and dust from the or receiver is free of dust and dirt. cover glass. (See page 3) Check that the target is within the Locate the target correctly in the measuring area. measuring area. (See page 9-4, 9-5) Check that the target is not leaned or Locate the target correctly in the shifted. measuring area. (See page 9-4, 9-5) The measured value displayed drifts.
Chapter 10 Troubleshooting Problem Inspection Remedy Check that the communications cable Connect the communications cable is correctly connected. correctly. (See page 8-2) RS-232C communications are not possible. The backlight does not turn ON. The image of the target does not appear. The remote control console does not operate. Check that the wiring specifications of the communications cable are correct. Check the wiring of the communications cable.
Chapter 10 Troubleshooting Error Messages The following section explains error messages that will be displayed if the LS-7600 Series has errors while in operation. Display Description Measuring head error ERROR-0 ERROR-1 This error message will be displayed if the connector to the measuring head is changed, the measuring head is not connected correctly, or there is a failure in the measuring head or connection cable. Calibration setting error Remedy Check the wiring of the connection cable.
Appendix This section provides information on optional accessories and the index of this manual. List of Optional Accessories .................................. Appendix-2 Setting Record Sheet ............................................ Appendix-3 Environment Setting Record Sheet ....................... Appendix-5 Index ...............................................................
Appendix List of Optional Accessories Item Type Appearance Explanation TR Cable (1m) OP-42182 A 1-m extension cable used to connect the transmitter and receiver. Up to two TR cables are connectable, provided that the total length is 6 m or less. TR Cable (3m) OP-42183 A 3-m extension cable used to connect the transmitter and receiver. Up to two TR cables are connectable, provided that the total length is 6 m or less.
Appendix Setting Record Sheet Use the following tables to record your setting details.
Appendix ■ Output Settings Item Default setting Setting detail OUT1 OUT2 OUT1 : +1A1 +0A2 Area calculation (CALC) OUT2 : +0A1 +1A2 Averaging number (AVE) Measuring mode (MEA-MODE) Offset Threshold 512 NORMAL 0.00000 Scaling at 10 V (A-OUT+10) +30.00000 Scaling at -10 V (A-OUT-10) Reference Scaling (A-SCALE) value Analog output setting: Analog update mode (A-SET) 1000μm/V Analog output setting: Output waiting voltage (A-CONT) +10.5V Comparator output setting: Strobe output time (STOBE-T) 0.
Appendix Environment Setting Record Sheet Use the following tables to record your environment setting details.
Appendix Index 1 One-point calibration ............................ 5-27 2 Two-point calibration ............................ 5-26 5 5-level comparator output ....................... 7-7 A A-TEST .................................................. 7-8 ALARM ................................................... 7-7 Alarm output ........................................... 7-7 Analog output settings.......................... 5-44 Analog scaling ............................. 5-44, 8-17 Analog voltage output ..
Appendix Focus monitor ............................... 5-9, 5-16 Function output selection ..................... 5-19 G Graph TIME ................................. 5-55, 5-62 Group processing ....... 3-9, 5-55, 5-63, 8-21 Group Processing Timing Chart ........... 7-14 H HH ........................................................ 8-19 HH set value ......................................... 5-50 HI .......................................................... 8-19 HI set value .................................
Appendix R Reference value mode ................ 5-51, 8-17 Remote control console ......................... 2-3 Reset .................................................... 8-11 RESET input .......................................... 7-7 Reset period and refresh cycle ............ 7-13 Response format .................................... 8-6 Rising edge .......................................... 5-10 RS-232C ......................................... 6-7, 8-2 RUN mode ...................................
WARRANTIES AND DISCLAIMERS (1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a period of one (1) year from the date of shipment. If any models or samples were shown to Buyer, such models or samples were used merely to illustrate the general type and quality of the Products and not to represent that the Products would necessarily conform to said models or samples.
Copyright (c) 2012 KEYENCE CORPORATION. All rights reserved.