User guide

7-15
Chapter 7 RS-232C
7
Measuring Mode
This command changes measuring mode settings.
LS
Output number
1 : OUT1
2 : OUT2
Measuring mode
00 : NORMAL 04 : AV-HOLD 08 : S-HOLD1
01 : P-HOLD 05 : A-P-HOLD 09 : S-HOLD2
02 : B-HOLD 06 : A-B-HOLD 10 : SELF-T
03 : D-HOLD 07 : A-D-HOLD
Command
Response
S D
,
M E
,
h k
,
S D
,
M E
,
h cr
crk
Self-timing Period
If the measuring mode is set to self-timing hold, this command will change self-timing
period settings.
LS
Output number
1 : OUT1
2 : OUT2
Command
Response
Self-timing period
0000~9999ms
crS D
,
T M
,
h l l l l
,
S D
,
T M
,
h cr
Offset
This command makes offset value setting changes.
LS
Output number
1 : OUT1
2 : OUT2
Command
Response
Value setting format
(See page 7-13)
S D
,
O F
,
h m m m m m m m m cr
,
S D
,
O F
,
h cr
Analog Scaling
This command changes scaling settings for analog output. The LS-7000 Series will
return an error response if a command received is different in tolerance mode.
Reference Value Mode
LS
Output number
1 : OUT1
2 : OUT2
Scaling value
00 : 1 03 : 10 06 : 100 09 : 1000 12 : 10000
01 : 2 04 : 20 07 : 200 10 : 2000
02 : 5 05 : 50 08 : 500 11 : 5000
Command
Response
nS D , A U , h ,
S D , A U , h cr
n cr