User guide

6-12
Chapter 6 I/O Terminals
6
Sample Hold 2
ON
OFF
TIMING input
ON
OFF
RESET/ZERO input
ON
OFF
P1 input to P4 input
ON
OFF
Comparator output
ON
OFF
WAITING output
ON
OFF
ON
OFF
STROBE output
BCD output
STROBE time
0±0.1ms
Min2.1ms
Max2.5ms
Min
2.1ms
Min2.1ms
Min2.1ms
Max2.5ms
Min 6.0ms
Min6.0ms
Max 8.0ms
Sampling
period
Triggered by the acknowledgment of timing input, the measured value will be deter-
mined with an elapse of the sampling period (i.e., reset time).
The strobe output time can be changed.
Refer to Strobe Output Time on page 4-48 for detail.
The sampling period (i.e., reset time) varies with the averaging number.
Refer to Reset Time and Refresh Cycle on page 6-13.
If RESET input or ZERO input is ON during the sampling period, the sampling will be
interrupted and the LS-7000 Series will await the result of measurement.
Self-timing Hold
YES
NO
Workpiece (Internal
measured value)
ON
OFF
RESET/ZERO input
ON
OFF
P1 input to P4 input
ON
OFF
Comparator output
ON
OFF
WAITING output
ON
OFF
ON
OFF
STROBE output
BCD output
STROBE time
0±0.1ms
Min0msMin
2.1ms
Min2.1ms
Min2.1ms
Max2.5ms
Sampling
period
ST
time
Max2.5ms
Min6.0ms
Min6.0ms
Max 8.0ms
Triggered by the acknowledgment of the workpiece, the measured value will be
determined with an elapse of self-timing and sampling periods (i.e., reset time).
The self-timing period is adjustable between 1 and 9,999 ms.
Refer to Self-timing Hold Setting on page 4-42 for detail.
The strobe output time can be changed.
Refer to Strobe Output Time on page 4-48 for detail.
The sampling period (i.e., reset time) varies with the averaging number.
Refer to Reset Time and Refresh Cycle on page 6-13.
If RESET input or ZERO input is ON during the sampling period, the sampling will be
interrupted and the LS-7000 Series will await the result of measurement.