Manual

Appendices
119
Appendices
ProblemItem
Inspection
Action
Ref. page
Measured
value
display
Displayed value is
incorrect.
Check if scanning head’s serial
No. matches that of ROM
board.
Check if calibration has been
performed correctly.
Check if optical axis is properly
aligned.
Check if target is tilted or
incorrectly positioned.
Check if target is correctly
placed in measurement area.
Check if reference edge is set
in measurement area.
Check target for dust or dirt.
Use scanning head with same
serial number as ROM board.
Perform calibration correctly.
Make sure that optical axis is
properly aligned in INSPECT
mode. If optical axis is mis-
aligned, adjust mounting
conditions while referring to
error message.
Place target correctly.
Place target in measurement
area correctly.
Set reference edge (when
T.EDGE/B.EDGE mode is
used).
Remove dust or dirt.
p.60
p.115
p.115
Analog
output
(Option)
No output
Check if measured value is
displayed.
Set controller and target correctly
so that measured value is dis-
played.
Output above +10 V
or below -10 V.
Check if analog scaling is set
correctly.
Make sure that analog scaling is
set correctly.
Computer
output
Computer data
output is disabled.
Check if connector pin connec-
tions are correct.
Check if computer has been
turned ON before controller.
Check if program is correct.
Connect connector pins cor-
rectly.
Turn ON computer first, and
then turn ON controller.
When " " is sent from compu-
ter, controller outputs measured
value.
p.8
p.57