User guide

4
Installation environment
To use this product properly and safely, avoid installing it in the following locations. Doing so
may lead to product breakdown.
Location that is humid, dusty or poorly ventilated
Location where the temperature becomes high, such as a place exposed to direct
sunlight
Location where there are flammable or corrosive gases
Location where the product may be directly subjected to vibration or impact
Location where water, oil or chemicals may splash onto the product
Considerations for improving noise immunity
Do not install in enclosures where high voltage devices are installed.
Locate as far away as possible from power lines.
Locate as far away as possible from devices that generate intense electric or magnetic
fields such as solenoids and chopper circuits.
Use separate conduit for I/O lines, power lines, and high voltage lines to prevent noise-
induced erroneous operation.
•Ground protective earth terminals and shield wires to a ground resistance of 100 ohms
or less.
Precautions on CE Marking
This device conforms to CE marking directives when the following conditions are met:
Make sure these conditions are met if using this device within the EU nations.
The applicable standards (EMC Directive) are listed below:
EMI:EN61326-1, Class A
EMS:EN61326-1
Limit the length of the power supply cable and all input/output cables that are connected to
the terminal panel of the controller to 30 m or less.