User guide

Table Of Contents
7-10
- IV Series User's Manual (Monitor) -
Clearing Errors
If an error targeted for clearing has occurred, outputting an error can be turned OFF by the clear error input.
Error output
Clear Error input
Min 2 ms
(Min 2 ms)
T5
(1)
(2)
ON
OFF
ON
OFF
Error occurred
(1) If an error occurred, the error output turns ON.
Error types which can be output are as follows.
System error
Startup memory readout error
External master registration error
Trigger error
To output the trigger error, set the trigger error in the Output Assignment (Page 4-60) under Settings
Navigator to [ON].
FTP error
To output the FTP error, set the FTP error within the FTP settings (Page 6-32) to [Enable].
(2) The error output can be turned OFF by the clear error input. Minimum ON/OFF input time is 2 ms.
Errors which can be cleared are as follows.
External master registration error
Trigger error
FTP error
The system error/startup memory readout error cannot be cleared.
For details on T5, refer to
“Input Response Time” (Page 7-12).
Clearing Errors
7
Controlling with the Input/Output Line