User guide
4-45
4
Setting Various Functions
IL-E
4-4 Calibration Function
When there is a difference between the internal measurement value (R.V.) and the actual 
dimension of the object, the value can be calibrated.
The zero point (offset) and tilt are both corrected.
When calculation function is disabled
Setting method (Calibrating R.V.)
Point
• If the following conditions are not satisfied, the screen below is 
displayed when the setting is complete. Make the setting again.
•
• Workpiece is not within measurement range.
• The setting was switched during the calibration.
Main screen
1
Press [MODE] button and X button for approx. 2 
seconds on the main screen at the same time. 
[30.CLb] is displayed on the main display (upper level).
2
Press 
S
 / 
T
 button to set [
USr
] and press 
X
 button.
Press [MODE] + 
X button
2 seconds or more 
at the same time
Calibration
R.V. you wish to display (2) [
5'V
]
R.V. before calibration (2)
R.V. you wish to display (1) [
5'V
]
R.V. before calibration (1)
Offset value A
Measurement value
Displayed value
'TT
LASER 
BANK
0 
1 
2 
3 
HI 
GO 
LO 
R.V. 
ANALOG 
HI SHIFT 
ZERO SHIFT 
TIMING 
LO 
5'6
HOLD CALC 
CHECK 
0.5 ≤
(R.V. you wish to display (2) ) - (R.V. you wish to display (1))
(R.V. before calibration (2)) - (R.V. before calibration (1))
≤ 2
LASER 
BANK
0 
1 
2 
3 
HI 
GO 
LO 
R.V. 
ANALOG 
HI SHIFT 
ZERO SHIFT 
TIMING 
LO 
HOLD CALC 
CHECK 
Setting 
value
Description
Default 
value
defLt Standard c
Usr User setting
*'#&
LASER 
BANK
0 
1 
2 
3 
HI 
GO 
LO 
R.V. 
ANALOG 
HI SHIFT 
ZERO SHIFT 
TIMING 
LO 
%#.KD
HOLD CALC 
CHECK 
Calibration setting