96M11843 Environment Resisting Type Built-in-amplifier Pressure Sensors GP-M Series Instruction Manual Read this instruction manual before using the product in order to achieve maximum performance. Keep this instruction manual in a safe place after reading it so that it can be used at any time.
Symbols In this instruction manual, the following symbols will be used to so that important points can be understood at one glance. Be sure to read these messages carefully. DANGER It indicates a hazardous situation which, if not avoided, will result in death or serious injury. WARNING It indicates a hazardous situation which, if not avoided, could result in death or serious injury. CAUTION It indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
CAUTION NOTICE 1. You must verify that the GP-M Series is operating correctly in terms of functionality and performance before the start and operation of the GP-M Series. 2. We recommend that you take all the necessary safety measures to avoid any damage in the unlikely event of a problem occurring. 3. Do not use the GP-M Series with corrosive liquids. 1.
CSA Certificate GP-M Series complies with the following CSA and UL standards and has been certified by CSA. • Applicable standard CAN/CSA C22.2 No.61010-1 UL61010-1 Be sure to consider the following specifications when using this product as a product certified by CSA. • Overvoltage category: 1 • Use this product under pollution degree 2. • Use this product at the altitude of 2000 m or less. • Indoor use only.
Precautions for wiring CAUTION NOTICE 1. Do not modify or disassemble unnecessarily. It may cause electric shock or improper operation. 2. Do not work on the wiring with bare hands. Even if the power is OFF, electric charge may remain and cause electric shock. 3. Do not work on the pipe work or wiring with wet hands. It may cause electric shock. 4. When wiring, confirm the pin position. 5. Use the GP-M Series within the rated range.
4. Do not remove the seal of the air hole of the GP-M001/M010/M025. It will no longer be waterproof. 5. When condensation occurs, it may become impossible to make measurements, or, damage may occur. Take measures such as the following to prevent condensation. NOTICE • Ensure that the ambient temperature is the same or lower than the fluid temperature. • Remove moisture via air conditioning • Separate the sensor 30 cm or more from the cooling pipes using the connector pipe. 6.
Name and Function of Each Part 4 Operating light (output 1) 5 Operating light (output 2) 1 Connector for input cable use 2 Clasp 7 DOWN 6 button (V) UP button (U) 8 MODE button * 3 Air hole * Only for the GP-M001/M010/M025 Piping/Installation Piping • Use the adapter for matching with the diameter of the piping. • When using a replacement adapter, attach the O-Ring included with the Main Unit to the groove of the G3/4 female screw part.
Precautions when installing z Attaching the coupling The recommended tightening torque when installing the adapter to the main body of the sensor is 20 N•m. It is recommended to apply grease to the G3/4 screw part in order to avoid scorching. z Grounding of metal parts The metal parts of the main body, the internal circuits 0 V are insulated.
Wiring, Output Diagram Output diagram z When choosing an NPN output Load 10 to 30 VDC 4 (Output 1) Load Main circuit Overcurrent protection circuit 1 2 (Output 2)* 3 0V * When choosing the function of output 2 QWV (control output) only. z When choosing a PNP output 10 to 30 VDC 2 (Output 2)* 3 Load 4 (Output 1) Load Main circuit Overcurrent protection circuit 1 0V * When choosing the function of output 2 QWV (control output) only.
Initial Settings at the Time of Delivery When turning the power ON for the first time after delivery, set the initial settings according to the outline below. * Alternate 2 P P2P MODE Push briefly Alternate K Q QWV MODE Push briefly Settings finish Polarity of control output Select by button. P2P NPN type 2P2 PNP type Choose function of output 2 Select by button. QWV Control output #P.
• The atmospheric pressure correction will be reflected on the analog output. Key Lock Function for preventing operation error. Activate/Deactivate Key Lock From the main screen, press and hold the MODE button and the U button (or the V button) simultaneously. If “.QE” is displayed while blinking, the Key Lock is enabled. With the same operation it can be disabled (“WP.” will be displayed while blinking).
Changing the Setting Value of Output 1 Change the setting value of output 1. The changing of setting values of output 2 is done by pressing and holding the MODE button. (Refer to page 15) Press or briefly Alternate 52 Switch-on point 1 Select using the buttons. Setting range: 1% of F.S. to 100% of F.S. (Details on page 17) *1. The detection mode of the Output 1 is *1 *2 MODE displayed as “(* ” when in window mode. Push briefly *2.
Reference • Peak (bottom) hold value will be stored into EEPROM every 45 seconds. • If the U and V buttons are pressed and held simultaneously on the peak (bottom) hold screen, the peak (bottom) hold value of the initial reset will be cleared. • To return to the previous screen, press the MODE button briefly.
Reference Pressing MODE and U at the same time will return to the previous screen. Initialization Value List Item Model Unit SP1/FH1 (Switch-on point 1) rP1/FL1 (Switch-off point 1) SP2/FH2 (Switch-on point 2) rP2/FL2 (Switch-off point 2) ASP (Analog start point (4 mA)) AEP (Analog end point (20 mA)) Default value (by model) GP-M GP-M GP-M GP-M GP-M GP-M 001 010 025 100 250 400 kPa MPa 40.0 0.400 1.000 4.00 10.00 16.00 20.0 0.200 0.500 2.00 5.00 8.00 100.0 1.000 2.500 10.00 25.00 40.00 80.0 0.
Settings of Each Setting Function Press and hold MODE button Alternate QW *PQ MODE Push briefly Output 2 at time of control output Alternate QW *PQ MODE Push briefly Alternate 52 *1 MODE 1. Configuration of output 1 Select using the buttons. *PQ Hysteresis Mode N.O. *P% Hysteresis Mode N.C. (PQ Window Mode N.O. (P% Window Mode N.C. (Details on page 17) 2. Configuration of output 2 Select using the buttons. *PQ Hysteresis Mode N.O. *P% Hysteresis Mode N.C. (PQ Window Mode N.O.
Alternate %Q( MODE Push briefly Alternate F5 MODE Push briefly Alternate FT MODE *2 Push briefly 7. Offset correction Select using the buttons. Setting range: Present value ±10% of F.S. (Details on page 19) MODE Push briefly Alternate FT MODE Push briefly Alternate 52'F MODE Push briefly MODE *3 Push briefly 8. ON delay timer of output 1 Select using the buttons. Setting range: 0.00 to 50.
Explanation of All Functions Detection mode and setting value The GP-M Series has two types of detection modes which are described below. When the selected function of output 2 is QWV (control output), the output 1 and output 2 can be set separately. Hysteresis mode (*PQ/*P%) Considering 5R T2 = hysteresis, it is the mode for changing the setting value when ON and when OFF. • When the output is OFF (right diagram 1), and the measured 52 value becomes more 52 than it will turn on (right diagram 2).
If *;5 is fixed at 0.5% of F.S., the presence or absence of hysterisis can be selected by the extended function (page 16). Reference The name of the setting values depending on the mode and output are as follows. Output Detection Mode Output 1 Output 2 Output 1 Output 2 when choosing hysteresis mode when choosing window mode Switch on Switch off point point 52 T2 52 * T2 * (* (. (. * (* * * Selection is possible when the selection function of output 2 (control output) is QWV.
Offset correction Concerning the value at the time of Display value Solid line: shipping, there may be as much condition at shipping as ±10%F.S. offset correction. This activates in the case of a margin of error in the displayed pressure value and the actual pressure value. ±10%F.S. Pressure value The pressure value after the offset correction is displayed on the setting screen and can be adjusted with the UV buttons.
Reference Time chart Output judgment ON OFF Control output ON after delay OFF F5 /F5 (ON delay) FT /FT (OFF delay) Responsiveness The time until 100% response of the internal judgment value can be set from within the range of 3 to 5000 (ms). The response time of the analog output (90% response) will be an additional 2 ms from this point. Power Save Mode When QP is selected, if the key operation does not occur for 15 seconds (Power Save) will be displayed.
Error Display and Countermeasures Error display 'T% Q. W. 'T2 'T' 'T5 Cause Measures At the time of atmospheric pressure correction there is ±10% F.S. applied pressure. There is too much current flowing through the control output. Outside possible display range (above) Outside possible display range (below) Return the air to the room pressure, once again set the sensitivity setting. Check the load, and return to the rated range. Return to the rated pressure range.
Specifications (Differs by model type) Model Rated pressure GP-M GP-M GP-M GP-M 001 010 025 100 -100 to -0.1 to -0.1 to 0 to +10 +100 +1 +2.5 MPa kPa MPa MPa -120.0 to -0.210 to -0.360 to -1.00 to +120.0 +1.110 +2.760 +11.00 kPa MPa MPa MPa GP-M 250 GP-M 400 0 to +25 0 to +40 MPa MPa -2.50 to -4.00 to Possible display +27.50 +44.00 range MPa MPa Allowable 400 kPa 4 MPa 10 MPa 30 MPa 50 MPa 50 MPa pressure 100 100 100 Burst pressure 1.5 MPa 15 MPa 35 MPa MPa MPa MPa Display 0.001 MPa 0.
Output Rating Medium temperature -20 to +100°C (no freezing/condensation)*3 *6 10-30 VDC, Ripple (P-P): 10% max, Power supply voltage Class 2 or LPS 50 mA or less (when 24 V: 32 mA or less, when Current consumption 12 V: 48 mA or less.
Material properties Pressure port: SUSXM7/ Wetted part Diaphragm pressure port: Al2O3/O-Ring: FKM Other parts Applicable cable Weight Housing metal portion: SUS304, SUS303 Housing plastic portion: PPSU Air hole*7: PTFE, nickel-plated brass. M12 connector 4 pin App. 150 g Relative humidity (%RH) *1 This is the value when considering linearity + hysterisis + repeatability in a stable environment of 23°C.
Dimensions (Unit : mm) ø34.2 44.6 15.4 2.6 29.2 44.3 86.7 M12 × 1 G3/4 A Useable length 10 * (Two surface width) 27 ø3 0 (27.5) * There is no air hole on the GP-M100/ M250/M400. Detecting element ø11 depth 1.2 (GP-M001/M010/M025) ø7 depth 2.9 (GP-M100/M250/M400) O-ring for GP-M001/M010/M025 use O-ring set for GP-M100/M250/M400 use ø15 2.5 ø2 A A ø1 ø8 A ø1 A 2. 8 9. 1.5 25 1.9 A-A A-A 2.
Warranties and Disclaimers (1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a period of one (1) year from the date of shipment. If any models or samples were shown to Buyer, such models or samples were used merely to illustrate the general type and quality of the Products and not to represent that the Products would necessarily conform to said models or samples.
INDIRECT, INCIDENTAL, PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY DAMAGES RESULTING FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS OF INFORMATION, LOSS OR INACCURACY OF DATA, LOSS OF PROFITS, LOSS OF SAVINGS, THE COST OF PROCUREMENT OF SUBSTITUTED GOODS, SERVICES OR TECHNOLOGIES, OR FOR ANY MATTER ARISING OUT OF OR IN CONNECTION WITH THE USE OR INABILITY TO USE THE PRODUCTS, EVEN IF KEYENCE OR ONE OF ITS AFFILIATED ENTITIES WAS ADVISED OF A POSSIBLE THIRD PARTY’S CLAIM FOR