4-3 Detection Settings (Func)
4
Settings for Advanced Functions
4-9
- Digital Fiber Sensor FS-N10 Series User's Manual -
DATUM1 mode
In DATUM1 mode, the light intensity received 
without a workpiece is always corrected to 
"100.0". The setting value is also corrected 
according to the correction amount, so that the 
ratio of setting value and received light intensity 
is kept constant, with the result of stable 
detection. The display of the setting value does 
not change.
The DATUM mode is effective in environments where received light intensity varies
gradually, for example where the sensor head is prone to contamination or large
temperature changes.
Sensitivity setting in DATUM1 mode
The sensitivity setting procedure 
below is an example of two-point 
calibration (where the received 
light intensity is "100.0" when a 
workpiece is absent and "20.0" 
when a workpiece is present) 
using the thrubeam/retro-reflective 
model.
1
Press the [SET] button with no work-
piece present.
2
Press the [SET] button with a work-
piece present.
The light intensity received in full-light 
receiving state is displayed as "100.0".
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
Received light intensity
Fixed proportion
"100.0"
"5.0"
(Initial value)
Setting
value
0
100.0
100.0
20.0
60.0
60.0
Calibration point (Step 1)
Calibration point (Step 2)
1
2
The DTM indicator
lights up.
Display changes
to "
100.0".