Product Manual
11
5. Let the kit sit for at least 24 hours to ensure the temperature within the kit has equilibrated and the
humidity inside the jars has reached the correct levels.
6. After the 24 hour equilibration period, open the 1st jar and while keeping the jar opening covered as
much as possible, quickly remove the Kestrel unit and start the rst calibration period by following the
instructions for your unit.
Kestrel 3000-3500 (Three-Button)
7. With the screen showing C1 press the center button to start a 60 minute countdown. The
screen will ash between C1 and the remaining number of minutes till the rst calibration
measurement is taken. Once complete, the screen will display C2.
Kestrel 3000 (Two-Button)
7. With the Kestrel turned on and in Relative Humidity Mode, press the On and Mode button
at the same time for approximately 6 seconds to start the RH Calibration routine. The screen
will display C1. Press Mode to start the 60 minute timer till the rst calibration measurement
is taken. The screen will ash between C1 and the preprogramed RH calibration value for
60 minutes till the rst calibration measurement is taken. Once complete, the screen will
display C2.
8. Return the Kestrel to the rst jar and after ensuring the lid is closed tightly, return the jars to the zipper
case and if using, return the zipper case to the insulated cooler. Make a note of what time the calibration
routine will end.
9. After the countdown for the rst calibration measurement has ended, remove the Kestrel from the rst
jar and make sure any salt or water is wiped o.