11 Instant help 
In case of an error in the program process, briefly turn off the balance and disconnect 
from power supply. The weighing process must then be restarted from the beginning. 
Fault 
Possible cause 
The displayed weight does not glow. 
  The balance is not switched on. 
  Batteries are inserted incorrectly or empty 
  No batteries inserted. 
The  displayed  weight  is  permanently 
changing 
  Draught/air movement 
  Table/floor vibrations 
  Weighing pan has contact with other 
objects. 
  Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
The  weighing  value  is  obviously 
wrong 
  The display of the balance is not at zero 
  Adjustment is no longer correct. 
  Great fluctuations in temperature. 
  The balance is on an uneven surface. 
  Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
Should other error messages occur, switch balance off and then on again. If the error 
message remains inform manufacturer. 