User manual
SFB-BA-e-1324 
22 
6.6.1  Verified models 
In verified weighing systems the menu item for adjustment „P2 mode“ is 
blocked. 
To override the blocked access you will have to destroy the seal before calling 
up the menu and to short-circuit the two contacts on the circuit board [K2], 
using a jumper (See chap.6.7). 
Attention: 
After destruction of the seal the weighing system must be re-verified by an 
authorised agency and a new verification wire/seal mark fitted before it can be 
reused for applications subject to verification. 
Call up menu: 
1.  Switch-on balance and during the selftest press  . 
2.  Press  ,  ,   subsequently, the first menu block 
„PO CHK“ will be displayed. 
3.   Press   repeatedly until „P2 mode“ will be displayed. 
4.   Press   and select the set weighing scales type by  . 
 = Single-range balance 
 = Dual range balance 
 = Multi-interval balance 
 
 
5.  Acknowledge with  . 
6.   Press   repeatedly until „CAL“ will be displayed. 
7.  Confirm with   and select setting „noLin“ by  . 