6.6  Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each display unit with connected weighing plate must be coordinated - in compliance 
with the underlying physical weighing principle - to the existing acceleration due to 
gravity at its place of location (only if the weighing system has not already been 
adjusted to the location in the factory). This adjustment process must be carried out 
for the first commissioning, after each change of location as well as in case of 
fluctuating environment temperature. To receive accurate measuring values it is also 
recommended to adjust the display unit periodically in weighing operation. 
  In weighing systems with a resolution of < 15 000 dividing steps an 
adjustment is recommended. 
In weighing systems with a resolution of > 15 000 dividing steps a 
linearisation is recommended (see chap. 6.6). 
  Prepare the required adjustment weight. The weight to be used 
depends on the capacity of the scale. Carry out adjustment as near as 
possible to the scale’s maximum weight. Info about test weights can be 
found on the Internet at: http://www.kern-sohn.com. 
  Observe stable environmental conditions. Stabilisation requires a 
certain warm-up time. 