6.3.2  Scope of delivery / serial accessories: 
  Balance, see chap. 2 
  Mains adapter 
  Protective cover 
  Instruction Manual 
6.4  Mains connection 
Power is supplied via the external mains adapter. The stated voltage value must be 
the same as the local voltage. 
Only use KERN original mains adapter. Using other makes requires consent by 
KERN. 
6.5  Battery power supply (option) 
The rechargeable battery is charged via the delivered power supply. 
Before the first use, the battery should be charged by connecting it to the mains 
power supply for at least 15 hours. 
  If an arrow appears on the weight display    above the battery symbol  , this 
is an indication that the capacity of the rechargeable battery will soon be 
exhausted. The balance will be ready to operate for about another 10 hours, then 
it will switch off automatically. Charge the battery with the help of the supplied 
power pack. 
  If in the weight display appears „bat Lo“ followed by a flickering display, the 
capacity of the rechargeable battery is below the prescribed minimum. The 
balance will be ready to operate for about another 5 minutes, then it will switch off 
automatically. Charge the battery with the help of the supplied power pack. 
The LED display informs you during loading about the loading status of the 
rechargeable battery. 
red: 
Battery is almost discharged 
green: 
Battery is completely discharged 
6.6  Initial Commissioning 
In order to obtain exact results with the electronic balances, your balance must have 
reached the operating temperature (see warming up time chap. During this warming 
up time the balance must be connected to the power supply (mains, accumulator or 
battery). 
The accuracy of the balance depends on the local acceleration of gravity. 
Strictly observe hints in chapter Adjustment. 