3 How to prepare the weighing balance
When the density set is installed, correct adjustment is not possible.
Perform necessary adjustment before installation of the density set using the
standard weighing plate.
If the balance is equipped with an internal adjustment weight (ALT / PLT, ALJ-A,
PLJ-A, PLJ-F) carry out an internal adjustment. After adjustment, enable “external
adjustment” in the adjustment mode menu. This will avoid a request of an internal
adjustment, when the density set is installed.
If the balance is equipped with an external adjustment weight (ALS-A, PLS-A, PLS-
F, PLE-N) carry out an external adjustment.
For carrying out please refer to the operating instructions supplied with each balance