User manual
PLJ-GM-BA-e-1312 
52 
9.5.3  Continuous output 
The balance may be operated in a mode enabling continuous output of weighing 
results. This mode can be turned on/off by commands via RS232. 
Data record format: 
1 
2 
3 
4 
5 
6 
7-15 
16 
17 
18 
19 
20 
21 
S 
I 
Space 
Stability 
indicator 
Space 
Pre 
sign 
Weig
ht 
Spac
e 
Unit 
CR 
LF 
Stability indicator  Space if stable,  
        ? if not stable 
        ^ if overload 
        v if underload 
Presign:     Space, if positive 
        negative sign, if negative 
Weight:      9 signs, right justified 
Unit:        3 signs, left justified 
10 Error messages 
„Err2“:        Value outside zero range 
„Err3“:        Value outside taring range 
„Err4“:    Calibration weight outside allowable range (+-1% for 
calibration weight) 
„Err5“:        Piece weight smaller than readability 
„Err7“:        Disconnection time was too short  
        (should be more than 3 seconds) 
„Err8“:        Impossible to carry out taring / resetting 
„NULL“:       Underload 
„FULL2“:      Weighing range exceeded 
„LH“:    Initial weight error Weight of weighing plate outside 
allowable tolerance of 10% 
„5-Full“:      Display range exceeded upon totalising 










