User manual
PLJ-GM-BA-e-1312
47
9 Data output RS 232 C
9.1 Technical data
7 / 8 data bits, 1 / 2 stop bit, no / even / odd parity bit
Baud rate selectable at 2400, 4800, 9600, 19200 and 38400 Baud
Miniature plug-in necessary (9 pole D-Sub)
For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)
Various transmission modes are available:
Manually after pressing
PRINT
Continuously, according to setting
Automatically according to stability display
Prompted by external device
(For remote control commands see chpt. 9.8)
9.2 Description of data output type
Operation:
Display:
Selecting the menu item
P2 Prnt
Press
TARE
, „2.2.Pr_n“ appears
2.1.Pr_n
Press
TARE
anew, the parameter selected as last
appears flashing
StAb
(Example)
Use and
TARE
to select data output type
rEPL CntA Cntb
StAb
Setting will be imported by pressing
PRINT
,
„2.1.PR_N“ appears
2.1.Pr_n
Press
F
repeatedly until “SAVE?“ appears
SAVE?
Acknowledge by
PRINT
, „0.0“
0.0 g