User manual

PLJ-GM-BA-e-1312
31
8.6.6 Charge display
This function can be used to suppress the charge function.
This suppresses the short illumination of the battery symbol if the balance is
operated without battery or accumulator. (supply by power pack)
The kind of supply is shown when the balance is started. SLA
CHr6 YES SLA Function activated / SLA battery in use
ATTENTION! Explosion hazard: For battery operation the setting must be no
Operation:
Display:
Selecting the menu item
P5 othr
Press
TARE
5.1.bl
Repeatedly press until the following display
appears
5.5. CHr6
To change your selection, press
TARE
.
no
To change your selection, press .
YES
Setting will be imported by pressing
PRINT
.
5.5.CHr6