9.5.3 Continuous output
The balance may be operated in a mode enabling continuous output of weighing
results. This mode can be turned on/off by commands via RS232.
Data record format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
Space
Stability
indicator
Space
Pre
sign
Weig
ht
Spac
e
Unit
CR
LF
Stability indicator Space if stable,
? if not stable
^ if overload
v if underload
Presign: Space, if positive
negative sign, if negative
Weight: 9 signs, right justified
Unit: 3 signs, left justified
10 Error messages
„Err2“: Value outside zero range
„Err3“: Value outside taring range
„Err4“: Calibration weight outside allowable range (+-1% for
calibration weight)
„Err5“: Piece weight smaller than readability
„Err7“: Disconnection time was too short
(should be more than 3 seconds)
„Err8“: Impossible to carry out taring / resetting
„NULL“: Underload
„FULL2“: Weighing range exceeded
„LH“: Initial weight error Weight of weighing plate outside
allowable tolerance of 10%
„5-Full“: Display range exceeded upon totalising