Data record format: 
Instru
ction: 
Space / 
instr.: 3. 
Character 
Stability 
indicator 
Space 
Presign 
Weight 
Space 
Unit 
CR 
LF 
Instruction: 1. up to 3 signs 
In case of a 3 digit instruction use the next space. This will not affect the length of the 
data record. 
Stability indicator    Space if stable,  
        ? if not stable 
        ^ if overload 
        v if underload 
Sign:        Space, if positive 
        negative sign, if negative 
Weight:        9 signs, right justified 
Unit:        3 signs, left justified 
9.5.2  Manual output 
The user may start output by pressing 
PRINT
. 
Settings in chapter 9.4 
Data record format: 
1 
2 
3 
4 -12 
13 
14 
15 
16 
17 
18 
Stability 
indicator 
Space 
Presign 
Weight 
Space 
Unit 
CR 
LF 
Stability indicator    Space if stable,  
        ? if not stable 
        ^ if overload 
        v if underload 
Sign:        Space, if positive 
        negative sign, if negative 
Weight:        9 signs, right justified 
Unit:        3 signs, left justified 