User manual
PLJ-GM-BA-e-1312 
47 
9   Data output RS 232 C 
9.1  Technical data 
  7 / 8 data bits, 1 / 2 stop bit, no / even / odd parity bit 
  Baud rate selectable at 2400, 4800, 9600, 19200 and 38400 Baud 
  Miniature plug-in necessary (9 pole D-Sub) 
  For operation  with interface faultless  operation  is only ensured with the  correct 
KERN – interface cable (max. 2m) 
Various transmission modes are available: 
  Manually after pressing 
PRINT
  Continuously, according to setting 
  Automatically according to stability display 
  Prompted by external device 
  (For remote control commands see chpt. 9.8) 
9.2  Description of data output type 
Operation: 
Display: 
  Selecting the menu item 
P2 Prnt 
   Press 
TARE
, „2.2.Pr_n“ appears 
2.1.Pr_n 
   Press 
TARE
 anew, the parameter selected as last 
appears flashing 
StAb 
(Example) 
  Use   and 
TARE
 to select data output type
rEPL  CntA  Cntb 
 StAb 
  Setting will be imported by pressing 
PRINT
, 
„2.1.PR_N“ appears 
2.1.Pr_n 
   Press 
F
 repeatedly until “SAVE?“ appears 
SAVE? 
  Acknowledge by 
PRINT
, „0.0“ 
0.0 g 