User manual
PLJ-GM-BA-e-1312 
14 
6.7  Verification 
General: 
According to the EU guideline 90/384/EEC balances must be verified officially if they 
are to be used as follows (legally regulated area): 
a)  For commercial transactions if the price of goods is determined by weighing 
b)  For the  production of  medines  in  pharmacies as well as for analyses in the 
medical and pharmaceutical laboratory 
c)  For official purposes 
d)  For the production of finished packages  
In case of doubt, please contact your local office of weights and measures.  
Verification information 
An EU qualification approval is available for those balances  marked as appropriate 
for  verification in the technical data. In  the event that the  balance is applied  in  an 
area subject to verification as described above, it must be verified and re-verified at 
regular intervals. 
Re-verification  of  a  balance  is  carried  out  in  compliance  with  the  respective  legal 
provisions of the states. The term of verification validity for balances in Germany, for 
example, is normally 2 years. 
The legal provisions of the country of use are to be observed.  
Position of the “official seals“: 
Balances with obligation to verify must be taken out of operation if: 
-  The weighing result of the balance is outside the error limit. Therefore, in 
regular intervals load balance with known test weight (ca. 1/3 of the max. load) and 
compare with displayed value. 
-  The reverification deadline has been exceeded. 
Position of the official seals 