  Put on weight 1 (see table chap. 6.8.1) 
  Wait for the stability display and the signal sound 
  „LoAd 2“ appears 
  Remove weight 1 and put on the weight 2 
  After the stability display and the signal sound 
„LoAd 3“ appears; remove weight 2 and put on 
weight 3 
  After the stability display and the signal sound 
„LoAd 4“ appears; remove weight 3 and put on 
weight 4 
  After having placed weight 4, „LoAd 0“ appears 
again 
  Remove weight 4, „LoAd 4“ appears anew 
  Place weight 4 again 
  Wait for the stability display and the signal 
sound, „LoAd 3“ will be displayed 
  Remove weight 4 and place weight 3 
  Wait for the stability display and the signal 
sound, „LoAd 2“ will be displayed 
  Remove weight 3 and place weight 2 
  Wait for the stability display and the signal 
sound, „LoAd 1“ will be displayed 
  Remove weight 2 and place weight 1 
  Wait for the stability display and the signal sound 
  „LoAd 0“ will be displayed 
  Remove weight 1 
  After the stability display and the signal sound 
the balance carries out a selftest and changes 
into the weighing mode. Now the linearization is 
concluded successfully. 