6.7.1  Procedure when adjusting (non-verifiable models): 
Observe stable environmental conditions. 
A warming up time (see chapter 1; Technical Data) is required for stabilization. 
Ensure that there are no objects on the weighing plate. 
Balance display 
Operation 
  In weighing mode press   approx. 3 sec. until 
„UnLoAd“ appears. 
When display „LoAd“ appears, put the necessary 
adjustment weight (see chap.1; Technical Data) on the 
center of the weighing plate. 
Adjustment will be carried out automatically after dead 
stop control. 
  Wait for stability display 
„Pass“ appears, the adjustment process has been 
finished successfully. 
   Take away adjustment weight 
Wait until the balance is again in the weighing mode. 