Operation 
  Start balance by pressing 
ON
OFF
. While the balance 
carries out a self test, press   until „F1 Unt“ 
appears 
   Press   repeatedly until “tECH“ appears 
In case of verifiable models press the adjustment switch on the lower side of the 
balance! 
   Press  , „Pin“ appears 
  Press  ,   and   successively, „P1 Lin“ 
appears 
   Press   anew, „Pin“ appears again 
  Press successively  ,   and  , „LoAd 0“ 
appears; after the apparition of the stability display 
and the signal sound, „LoAd 1“ will be displayed 
  Put on weight 1 (see table chap.6.8.1) 
  Wait for the stability display and the signal sound 
  „LoAd 2“ appears 
  Remove weight 1 and put on the weight 2 
  After the stability display and the signal sound „LoAd 
3“ appears; remove weight 2 and put on weight 3 
  After the stability display and the signal sound the 
balance carries out a selftest. Remove the weight 
during selftest. The balance changes into weighing 
mode. Now the linearization is concluded 
successfully. 