User manual
16    PFB-BA-e-1322 
„Pass“ appears, the adjustment process has been 
finished successfully. 
   Remove adjusting weight during selftest 
Wait until the balance is again in weighing mode. 
6.8  Linearization 
Linearity shows the greatest deviation of a weight display on the scale to the value of 
the  respective  test weight  according  to  plus  and  minus  over  the  entire  weighing 
range. 
If  linearity  deviation  is  discovered  during  a  testing  instrument  control,  you  can 
improve this by means of linearization. 
  In balances with a resolution of > 15 000 dividing steps carrying out a 
linearization is recommended. 
  Carrying out linearization is restricted to specialist staff possessing well 
acquainted with the workings of weighing scales. 
  The test weights to be used must be adapted to the weighing scale’s 
specifications; see chapter “testing instruments control”. 
  Observe stable environmental conditions. Stabilisation requires a certain 
warm-up time. 
  After successful linearization you will have to carry out calibration;  
see chapter “testing instruments control”. 