4 Operation
1. Connect mains adapter and start the balance by pressing the ON/OFF key
2. Entering limiting values
3. Evaluation of readings via relay contacts
4. Status balance/contact
Status balance Display Contact
Reading outside condition No reading display inactive
Reading outside range No reading display inactive
Input of invalid values Five symbols appear
on the left
inactive
Error message
(o-ERR/u-ERR)
o-ERR/u-ERR
OFF during measure-
ment of limits 1-5
ON during ERR
During input Input appears on display inactive
Waiting for stability
message as per tares
[M] flashing Last display fixated
Stand-by Stand by LED glowing inactive
Sleep mode Sleep LED glowing Output of evaluation result
Automatic check of
balance at starting
[88888888] glowing
inactive
5. Output by external signal
Contacts and audio signals are started by the weighing result, but may also be
triggered by an external signal
Whilst the menu is set [2B. R.Q.m. 2] there is an output of the weighing result for
approx. 200 ms after the signal was applied
For pulsed, external signals the output signal will follow after 100 ms at the
earliest