User manual
PES/PEJ-BA-e-1316 
58 
There are two different ways to set the tolerance limits: 
1.  Place the values (object) on the balance - 
          > Save this value 
2.  Numeric entering of values - 
          > Enter the limits via keyboard. 
NOTE: 
  If a limiting value was set it remains saved until the balance is turned off. 
  For the functions weighing, counting, percent individual limits can be set. 
  When entering the limits please pay attention to the type of evaluation that was set. 
12.2.  Display of the results 
12.2.1.  For 2 limits 
The triangular tolerance marker (
) in the upper part of the display shows whether 
the goods to be weighed are within the two tolerance limits. 
The tolerance marker is only in operation during operating mode tolerance weighing; 
it is otherwise not visible. 
The tolerance marker provides the following information: 
Goods to be weighed above tolerance limit 
Goods to be weighed within tolerance range 
Goods to be weighed below tolerance limit 
Display 
Result 
If a point is set as lower 
limit 
If two points are set as upper and 
lower limit 
+   (high) 
No display 
Weight > Upper limit 
TOL  (OK) 
Lower limit   Weight 
Lower limit  Weight   Upper limit 
-   (low) 
Lower limit > Weight 
Lower limit > Weight 