User manual
PES/PEJ-BA-e-1316 
24 
6.9.4.  Adjustment with internal weight (only PEJ) 
With the internal adjustment weight, the weighing accuracy can be checked and re-
adjusted at any time. 
Procedure when adjusting: 
Observe stable environmental conditions. A warming up time of ca. 1 hour is re-
quired for stabilization. Ensure that there are no objects on the weighing plate. 
Operation 
Display 
Activate function [   ] 
(see chap. 7). 
 
g
Start of the automatic adjustment: 
Adjustment is carried out automatically. 
The process of adjustment is completed. 
The balance returns automatically into 
weighing mode. 
CAL
 
 
CAL
 
CAL
 
 