KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instruction Precision balance KERN PES/PEJ Version 1.
GB KERN PES/PEJ Version 1.6 04/2013 Operating instruction Precision balance Contents 1. Technical data ................................................................................................ 5 2. Declaration of conformity ........................................................................... 10 3. Basic Information (General) ........................................................................ 11 3.1. Proper use ......................................................................
6.9.1. Adjustment with external weight (only PES) ......................................... 19 6.9.2. Adjustment test with external weight (only PES) .................................. 21 6.9.3. Automatic adjustment (only PEJ) ........................................................... 23 6.9.4. Adjustment with internal weight (only PEJ) ........................................... 24 6.9.5. Adjustment test with internal weight (only PEJ) ................................... 25 6.10. Verification .
12.4.1. Entering 2 limits by weighing .................................................................. 60 12.4.2. Entering 3 or 4 limits by weighing .......................................................... 63 12.4.3. Numeric entering of 2 limits .................................................................... 66 12.5. Evaluation with difference values ........................................................... 69 12.5.1. Entering 2 limits by weighing ...............................................
1. Technical data KERN PES 220-3M PES 420-3M PES 620-3M 0.001 g 0.001 g 0.001 g Weighing range (max) 220 g 420 g 620 g Minimum load (Min) 0.02 g 0.02 g 0.1 g Verification value (e) 0.01 g 0.01 g 0.01 g Accuracy class II II I Reproducibility 0.001 g 0.001 g 0.001 g ± 0.002 g ± 0.003 g ± 0.003 g 3 sec. 3 sec. 3 sec.
KERN PES 2200-2M PES 4200-2M PES 6200-2M Readability (d) 0.01 g 0.01 g 0.01 g Weighing range (max) 2,200 g 4,200 g 6,200 g Minimum load (Min) 0.5 g 0.5 g 1g Verification value (e) 0.1 g 0.1 g 0.1 g Accuracy class II II I Reproducibility 0.01 g 0.01 g 0.01 g ± 0.02 g ± 0.02 g ± 0.03 g 3 sec. 3 sec. 3 sec.
KERN PES 8200-1M PES 15000-1M PES 31000-1M 0.1 g 0.1 g 0.1 g 8,200 g 15,000 g 31,000 g Minimum load (Min) 5g 5g 5g Verification value (e) 1g 1g 1g Accuracy class II II II Reproducibility 0.1 g 0.1 g 0.1 g Linearity ± 0.2 g ± 0.2 g ± 0.4 g Stabilization time 3 sec. 3 sec. 3sec. 5 kg + 2 kg (F1) 10 kg + 5 kg (F1) 20 kg+10 kg(F1) 4 4 8,9 0.1 g 0.1g 0.
KERN PEJ 220-3M PEJ 420-3M PEJ 620-3M 0.001 g 0.00 1g 0.001 g Weighing range (max) 220 g 420 g 620 g Minimum load (Min) 0.02 g 0.02 g 0.1 g Verification value (e) 0.01 g 0.01 g 0.01 g Accuracy class II II I Reproducibility 0.001 g 0.001 g 0.001 g ± 0.002 g ± 0.003 g ± 0.003 g 3 sec. 3 sec. 3 sec.
KERN PEJ 2200-2M PEJ 4200-2M Readability (d) 0.01 g 0.01 g Weighing range (max) 2,200 g 4,200 g Minimum load (Min) 0.5 g 0.5 g Verification value (e) 0.1 g 0.1 g Accuracy class II II Reproducibility 0.01 g 0.01 g ± 0.02 g ± 0.
2. Declaration of conformity KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.
3. Basic Information (General) 3.1. Proper use The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read. 3.2. Improper Use Do not use balance for dynamic weighings.
3.4. Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this.
6. Unpacking, Setup and Commissioning 6.1. Installation Site, Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Your balance in overview: Air bubble Covering lid over underfloor weighing device Rear view: RS 232 interface Printer interface Connection net adapter 14 Antitheft device PES/PEJ-BA-e-1316
6.3. Scope of delivery Serial accessories: Balance Weighing plate Mains power supply Operating Manual Protective cover 6.4. Setup Positioning of weighing plate: Levelling balance: Use the foot screws to level the balance until the air bubble in the bubble level is in the prescribed circle.
6.5. Mains connection Power is supplied via the external mains power supply. The stated voltage value must be the same as the local voltage. Only use original KERN mains power supplies. Using other makes requires consent by KERN. In the menu you can activate the AUTO-SLEEP function . In net operation the balance after 3 min without load change or key pressure passes in a sleep mode. Automatic activation of the display by load change or by pressing any key. 6.6.
6.8. Initial Commissioning A warming up time of 10 minutes after switching on stabilizes the measuring values. The accuracy of the balance depends on the local acceleration of gravity. Please be sure to observe the information in the chapter on adjusting in chap. 6.9. 6.8.1. Power display Supply balance with power via the mains power supply. Balance is in stand-by mode (green LED is on). Use the ance on.
6.8.2. Bar graph display In configuration menu 1 (chapter 7) you can activate/deactivate the bar graph display. * weihging dish ist unloaded Wägeplatte unbelastet Hide bar graph Show bar graph half of weighing range Waage weighing range is used halbe Höchstlast ist mit Höchstlast belastet to full capacity The weighing range of the balance is divided into 40 graphic cuboids. Zero (0) will appear on the graphic display if there is no weighing value on the balance.
6.9. Adjustment As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out at every commissioning, after each change of location as well as in case of fluctuating environment temperature.
Operation Activate function [ Display ] (see chap. 7). g CAL CAL Zero point will be saved. CAL Carefully place adjusting weight in the centre of the weighing plate CAL Adjustment process is started. CAL The process of adjustment is completed. Remove adjusting weight, balance will return into weighing mode automatically. In case of an adjustment error or incorrect adjusting weight the display will show [- Err]; repeat adjustment process.
6.9.2. Adjustment test with external weight (only PES) During adjustment tests the balance automatically compares the saved value of the adjustment weight with the actual value. This is only a check, i.e. no values are changed. Procedure: Observe stable environmental conditions. A warming up time of ca. 1 hour is required for stabilization. Ensure that there are no objects on the weighing plate. Operation Activate function [ (see chap. 7).
Start of the adjustment test: CAL CAL CAL Carefully place adjusting weight in the centre of the weighing plate CAL The difference between the saved value and the measured value is displayed. g Take away adjustment weight. Press any key; the adjustment process is cancelled and the balance returns to weighing mode.
6.9.3. Automatic adjustment (only PEJ) After switching on the balance, the automatic adjustment starts with the internal adjustment weight. Switch on the balance using ON/OFF. The balance carries out a selftest, „M“ will be displayed flashing. After that the automatic adjustment starts: „Aut. CAL“ appears flashing, followed by „CH. 0“ and „CH. F.S.“ If „End“ is displayed, the automatic adjustment has been concluded successfully.
6.9.4. Adjustment with internal weight (only PEJ) With the internal adjustment weight, the weighing accuracy can be checked and readjusted at any time. Procedure when adjusting: Observe stable environmental conditions. A warming up time of ca. 1 hour is required for stabilization. Ensure that there are no objects on the weighing plate. Operation Activate function [ (see chap. 7). Display ] g Start of the automatic adjustment: CAL Adjustment is carried out automatically.
CAL g 6.9.5. Adjustment test with internal weight (only PEJ) During adjustment tests the balance automatically compares the saved value of the adjustment weight with the actual value. This is only a check, i.e. no values are changed. Procedure: Observe stable environmental conditions. A warming up time of ca. 1 hour is required for stabilization. Ensure that there are no objects on the weighing plate. Operation Activate function [ (see chap. 7).
CAL The difference between the saved value and the measured value is displayed. g Press any key; the adjustment process is cancelled and the balance returns to weighing mode.
6.10. Verification General introduction: According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legally controlled area): a) For commercial transactions if the price of goods is determined by weighing b) For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory c) For official purposes. d) For manufacturing final packages. In cases of doubt, please contact your local trade in standard.
6.10.1. Seals and unlocking switch Position: Unlocking switch Seal Seal After verification the balance is sealed at the indicated positions. Verification of the balance is invalid without the "seal". Access to the unlocking switch by removing the seal (verification will be invalid!) and the rubber plug (see drawing).
7. Application and configuration menu 1 In the menu the settings of the balance can be modified and functions can be activated. This way, the balance can be adjusted to individual weighing requirements. The menu is structured as follows Application menu: To adjust the balance to user requirements Configuration menu 1: Definition of the basic functions 7.1. User principle of the menu control Operation Display Switch on balance: g Call up menu: Press for about 4 seconds, until [ ] is displayed.
Changing the function: or Further pressing the keys will call up the various functions of the menu (see table chapter 7.2) Changing the parameter: or To change the last digit of the parameter, actuate the TARE-key or the arrow keys. Saving your settings: g Leave the function menu and return to weighing mode General information about using the arrow keys for entering: Operation via arrow keys is faster and more comfortable than via the TARE and F key.
7.2. Menu overview The manufacturer’s setting has a certain standard configuration. This one is marked with *. Function Display or Selection or options Measure medium * Data output * Autom. Data output * Weighing Parts counting Percent determination Density determinination of solids dest. Water Measuring liquid of your selection only output measuring value density Output of all density parameters OFF (Output only after pressing PRINT key) * ON Off * Adding Tolerance weighing see chpt. 7.2.
Adjustment * : Factory setting PEJ * : Factory setting for Class I PES * : Factory setting for Class II PES Bar graph Automatic turn-off for battery operation (function only exists for battery operation) * * * * * Auto Sleep-Funktion in mains operation * Units A * CAL-key deactivated Automatic internal adjustment Adjustment with external weight External adjustment Adjustment test with external weight Hide bar graph Show bar graph Automatic turn-off after 3 min.
7.2.1. Parameter additional functions Not displayed at menu setting „ Function Display “ Selection or or Display conditions of the tolerance marker * Tolerance range * Number of limiting points Description of options * Tolerance marker is always displayed, even if standstill control is not yet displayed. Tolerance marker is only displayed in connection with standstill control. Tolerance marker is only displayed above zero range (mind + 5). Tolerance marker is displayed for the whole range.
7.2.2. Parameter for serial interface Not shown for menu setting „ Function Display or “ (interface de-activated). Selection Description of the or options No data output Continuous data output Continuous data output stable weighing values Output for stable and instable weighing values after pressing PRINT key Output for stable weighing value after previous relief of balance One output for stable weighing value. No output for stable weighing values.
Parity only at setting or * No parity bit Odd parity Even parity Data Bits 7 bits only at setting * Stop Bits 8 bits 1 bit only at setting * 2 bit * Always use this setting * Always use this setting not documented not documented PES/PEJ-BA-e-1316 35
8. Configuration menu 2 8.1. User principle of the menu control Operation Display Switch on balance: g Call up menu: + Hold F-key while TARE key is pressed until [ ] is displayed. When releasing, the first function is displayed [ ] Changing the function: or Further pressing the keys will take you through the various functions of the menu.
Changing the parameter: or To change the last digit of the parameter actuate TARE-key or arrow keys. Saving your settings: Leave the menu and return to weighing mode g 8.2. Menu overview The manufacturer’s setting has a certain standard configuration. This one is marked with *. Function Display Choice * Description of the options Off Setup balance ID no.
9. Operation 9.1.
9.2. Overview of display Display Description g, kg Gram, Kilogram 0 Zeroing display - Minus ο Stability display Net Tara symbol B/G Gross weight Pcs Parts counting % Percent weighing Tolerance weighing * Adding function active ∑ Total L Output date/time M CAL Balance carries out balance function, e.g. unit count / display of stored value Display for adjustment. Signals the adjustment function. Weighing unit display Bar graph Message for battery mode (optional) see chpt. 6.
10. Weighing mode This way, 4 different weighing modes are available for you: 1. Weighing [ ] 2. Weighing/parts counting [ ] 3. Weighing/percent determination [ ] 4. Weighing/density determination [ ] Excepted weighing / density determination you can activate, with the selection of of a weighing mode, more functions like e.g. tolerance weighing, adding (see chap. 7.2 „Additional functions“). So you can display the measuring values according to your needs. Actuating the "Pcs").
10.1.1. Taring The dead weight of any weighing container may be tared away by pressing a button, so that the following weighings show the net weight of the goods to be weighed. Operation Display Place empty tare container on the weighing plate. The total weight of the container is displayed. g Reset display to "0": Net g The weight of the container is now saved internally; in addition the display shows the tare symbol "Net". Place the goods to be weighed into the tare container.
The taring process can be repeated any number of times, e.g. when adding several components for a mixture (adding). Reset display to "0": Net g The total weight of the container is tared away. Add more components into the weighing container (adding). Net g Now read off the weight of the added item to be weighed on the display. NOTE: The balance is able to only store one taring value at a time. When the balance is unloaded the saved taring value is displayed with negative sign.
10.1.2. Net/gross The dead weight of any weighing container may be tared away by pressing a button. For subsequent weighings the net weight of the goods to be weighed as well as the gross weight goods + taring container can be displayed. Condition: * Function [ ] active (see chapter 7) Operation Display Place empty tare container on the weighing plate. The total weight of the container is displayed.
Net g Use the F key to switch from net weight to gross weight or vice versa This procedure may be repeated any number of times (max. weighing range of the balance). B/G g 10.2. Parts counting With parts counting you can either count parts into a container or remove parts from a container. To count a greater number of parts the average weight per part has to be determined with a small quantity (reference quantity). The larger the reference quantity, the higher the counting exactness.
Determine the reference unit: The display shows flashing the last saved reference quantity. Pcs Press for about 4 seconds until [U. SEt.] appears, then release M The display e.g. 10 Pcs. prompts you to enter 10 pieces as reference. Change reference quantity: or Use the TARE-key or the arrow keys to switch between the following reference quantities: Pcs M 10 5 30 100 Important: The larger the reference quantity, the more accurate the parts counting.
The reference weight is saved. Remove reference weight. Pcs Count the items: Now you can fill the items to be counted into the container. The respective quantity is shown in the display. By repeated pressing, switching option of the displayed value of the displayed value e.g.
10.3. Percent determination Percent weighing allows to display weight in percent, in relation to a reference weight. The displayed weighing value is stored as a standard percent value (default setting: 100%). 10.3.1. Entering the reference weight by weighing Operation Activate function [ (see chap. 7). Display ] % The display shows the %-symbol.
From now, the added weight is shown in %. % g By repeated pressing, switching option of the displayed value in „g“ or %“ % NOTE: If the error message “o-Err “ is displayed, the reference weight is outside the weighing range The 100% reference is preserved until the balance is disconnected from the mains. 10.3.2. Numeric entering of the reference weight Operation Activate function [ (see chap. 7). Display ] % The display shows the %-symbol.
g With "0" flashing, you are prompted to enter the reference weight numerically Entering the numeric value: or Þ Þ Þ » Þ Þ Þ Any time you press TARE-key or the arrow key, you will go through the numbers 0-9, decimal dot and minus Select the number to be changed (the active position flashes): or An acoustic signal sounds; the entered reference weight is saved PES/PEJ-BA-e-1316 % 49
From now, the added weight is shown in %. % g By repeated pressing, switching option of the displayed value in „g“ or %“ % NOTE: If the error message “o-Err “ is displayed, the reference weight is outside the weighing range The 100% reference is preserved until the balance is disconnected from the mains.
10.4. Density determination of solids (hydrostatic weighing) Density is the relationship of weight [g] : volume [cm 3]. The weight is determined by weighing the sample in air. The volume results from the ascending force [g] of the sample dipped in a liquid. The density [g/cm 3] of that liquid is known (principle of Archimedes). Operation Display The density is determined with help of the underfloor weighing device.
Input of temperature: or Þ Þ Þ » Þ Þ Þ Any time you press TARE-key or the arrow key, you will go through the numbers 0-9, decimal dot and minus Select the number to be changed (the active position flashes): or Save, an acoustic signal sounds g Once you have selected a measuring liquid according to your selection its density is input (input range 0.0001 to 9.9999 g/cm 3).
After the parameter input for the measuring liquid, the density of your sample will be determined 1. Weight of the sample in air g Tare balance with sample holder Put-on sample g Wait till the weight display on the balance is stable The weight of the sample in air is saved 2.
Place sample and immerse it ƒ g u Wait till the weight display on the balance is stable The density of your sample is displayed, characterized by the u symbol right above.
11. Adding of displayed values Any number or individual weighings are automatically added to a total, e.g. all individual weighings of a batch. The adding function is possible in all functions of the weighing mode (Weighing/parts counting/ percent determination). Operation 1. Activate function [ (see chap. 7). Display ] 2. Select one of the following settings [ ] : Add [ ] : Adding with AUTO-TARA 3. Place the weight A, wait until the stability display [] appears g 4.
7. Wait until stability display is shown [] : g The displayed value is added into the total adding memory. The total [ ] is briefly shown g g Remove weight and place further weights on balance; for each weight, repeat step 4 to 6 g 8. Total of all individual weighings: g By repeatedly pressing the F-key, switching of the displayed value into other activated functions. g 9. Delete total added memory: Display total (step 7), then press TARE-key. g 11.1.
12. Weighing with tolerance range 12.1. General Information This balance can be used as dispensing as well as sorting balance; the respective lower tolerance limit as well as upper tolerance limit can be programmed. An acoustic signal supports portioning, dispensing or sorting.
There are two different ways to set the tolerance limits: 1. Place the values (object) on the balance > Save this value 2. Numeric entering of values > Enter the limits via keyboard. NOTE: If a limiting value was set it remains saved until the balance is turned off. For the functions weighing, counting, percent individual limits can be set. When entering the limits please pay attention to the type of evaluation that was set. 12.2. Display of the results 12.2.1.
12.2.2. For 3 or 4 limits Display of tolerance mark + TOL Limit 5 Limit 4 Limit 3 Limit 2 Limit 1 Limit 5 4. Limit point Weight Limit 4 3. Limit point Weight < 4. Limit point Limit 3 2. Limit point Weight < 3. Limit point Limit 2 1. Limit point Weight < 2. Limit point Limit 1 Weight < 1. Limit point 12.3. Basic settings for weighings with tolerance range Operation Display 1. Activate tolerance weighing function [ ] or [ ] (see chap. 7). 2.
12.4. Evaluation of absolute values 12.4.1. Entering 2 limits by weighing Important information! Always begin by entering the lower limit value, followed by the upper limit value. Enter. Operation Display 1. Activate tolerance weighing function [ ] or [ ] (see chap. 7). 2. Actuate required parameter selection Parameter selection for 2 limiting points: or until [ ] or [ ] is displayed; more settings of your choice (see chap. 12.
4. Entering limiting values: Press for about 4 seconds, until [ ] is displayed, then release The flashing display (last saved value) prompts you to enter the lower limiting value ( ) 5. Place sample for the lower (i.e. smaller) limiting value on the weighing plate: 6. Save: An acoustic signal sounds, the saved lower limit is briefly displayed.
7. Place sample for the upper (i.e. larger) limiting value on the weighing plate: 8. Save: An acoustic signal sounds, the saved upper limit is briefly displayed. The balance returns to tolerance weighing mode. From here evaluation takes place whether the goods to be weighed are within the two tolerance limits. * If you want to set for your tolerance weighing only one limit point (parameter selection [ ]), ignore step 7 and 8.
12.4.2. Entering 3 or 4 limits by weighing Operation Display 1. Activate tolerance weighing function [ ] or [ ] (see chap. 7). 2. Actuate required parameter selection Parameter selection for 3 limiting points: or until [ ] or [ ] is displayed; more settings of your choice (see chap. 12.3) are carried out in the same way Parameter selection for 4 limiting points: Parameter selection for absolute value: 3.
4. Entering limiting values: Press for about 4 seconds, until [ ] is displayed, then release g The flashing display (last saved value) prompts you to enter the first lower limiting value ( ) 5. Place sample for the first limiting value on the weighing plate: 6. Save: An acoustic signal sounds, the saved first limiting value is briefly displayed.* g The flashing display (last saved value) prompts you to enter the second limiting value ( ) 7.
8. Save: An acoustic signal sounds, the saved second weighing value is briefly displayed. g The flashing display (last saved value) prompts you to enter the third limiting value ( ) 9. To enter 3rd and 4th limiting value, repeat steps 7 and 8 10. Save: An acoustic signal sounds, the last saved 3rd or 4th limiting value is briefly displayed. The balance returns to tolerance weighing mode. From here, evaluation is carried out whether the goods to weighed are within the tolerance limits.
12.4.3. Numeric entering of 2 limits Operation Display 1. Activate tolerance weighing function [ ] or [ ] (see chap. 7). 2. Actuate required parameter selection Parameter selection for 2 limiting points: or until [ ] or [ ] is displayed; more settings of your choice (see chap. 12.3) are carried out in the same way Parameter selection for absolute value: 3. Leave function menu g The balance is now in tolerance weighing mode; the tolerance marker ( ) appears 4.
Display changes to flashing "zero" 5. The flashing display prompts you to numerically enter the lower limit 6.
7. Save: An acoustic signal sounds, the saved lower weighing value is briefly displayed. The flashing display (last saved value) prompts you to enter the upper limiting value 8. To enter the numeric value for the upper limiting value, repeat steps 5 6 9. Save: An acoustic signal sounds, the saved upper limit is briefly displayed. The balance returns to tolerance weighing mode. From here evaluation takes place whether the goods to be weighed are within the two tolerance limits.
12.5. Evaluation with difference values 12.5.1. Entering 2 limits by weighing Important information! Always begin by entering the lower limit value, followed by the upper limit value. Enter. Operation Display 1. Tolerance weighing function [ ] or [ ] (see chapter 7). 2. Actuate required parameter selection Parameter selection for 2 limiting points: or until [ ] or [ ] appears; more settings of your choice (see chap. 12.3) are carried out in the same way Parameter selection for difference value: 3.
4. Entering a reference value: Press for about 4 seconds, until [ ] appears, then release. The flashing display (last saved value) prompts you to enter a reference value 5. Place reference weight onto weighing plate: 6. Save An acoustic signal sounds, the saved reference value is briefly displayed.
7. Place sample for the first limiting value on the weighing plate: 8. Save An acoustic signal sounds, the saved lower difference value is briefly displayed. The flashing display (last saved value) prompts you to enter the upper limiting value 9. Place sample for the upper (i.e. larger) limiting value on the weighing plate: 10. Save An acoustic signal sounds, the saved upper difference value is briefly displayed. Remove sample from weighing plate. The balance returns to tolerance weighing mode.
12.5.2. Entering 3 or 4 limits by weighing To enter 3 or 4 limiting values [ 8 (see also chapter 12.4.2). ]-[ ] or [ ], repeat steps 7 and Display of tolerance mark + TOL [ [ [ [ [ 12.5.3. ] ] ] ] ] 4. limiting point 3. limiting point Reference weight 2. limiting point 1. limiting point Numeric entering of 2 limits Operation Display 1. Activate tolerance weighing function [ ] or [ ] (see chap. 7). 2.
4. Entering a reference value: Press for about 4 seconds, until [ ] appears, then release. The last saved reference weight appears flashing Display changes to flashing "zero" 5. The flashing display prompts you to numerically enter a reference weight 6.
7. Confirm An acoustic signal sounds, the saved reference weight is briefly displayed. The flashing display (last saved value) prompts you to enter the lower difference value 8. Entering lower limit Repeat steps 5 and 6 9. Confirm An acoustic signal sounds, the saved lower difference value is briefly displayed. 10. The flashing display (last saved value) prompts you to enter the upper difference value 11.
11. Save An acoustic signal sounds, the saved upper difference value is briefly displayed. The balance returns to tolerance weighing mode. From here evaluation takes place whether the goods to be weighed are within the two tolerance limits. To enter 3 or 4 limiting values [ 9 (see also chapter 12.4.2). ]-[ ] or [ ], repeat steps 8 and 13. Setting date and time Display symbol [ L ] 13.1. Time Operation Display 1. Call up menu Keep pressed until [ ] appears.
Press again L : : Last stored time appears.* 2. Resetting time L : : Time to be changed is flashing Select the number to be changed (the active position flashes): L : : or Changing the numerical value L or : : After storing your setting, the date display will appear 3. Save L 4. Return to weighing mode g *Notice: 76 The display can be rounded up (from 30 s) or down (to 29 s) by pressing the TARE-key.
13.2. Date You can set the display of your data output under menu item overview chpt. 7.2.). Operation (see menu Display 1. Call up menu Keep pressed until [ ] appears.
2. Resetting the date L Time to be changed is flashing Select the number to be changed (the active position flashes): L or Changing the numerical value L or 3. Save After storing your settings, the balance will automatically return to weighing mode.
13.3. Interval output function This menu item allows you to determine after which interval you wish data output to be carried out. To achieve this, activate the [ the menu (see chpt. 7.2.1) 13.3.1. ] or [ ] function in Interval setting Operation Display 1. Call up menu [ keep pressed until ] appears. : : Time to be changed is flashing 2. Set interval Select the number to be changed (the active position flashes): : : : : or Changing the numerical value or 3.
13.3.2. Start/Stop interval output Operation Display Start output L g Stop output g The balance returns automatically into weighing mode.
13.4. Input balance ID-no. Display symbol [ ] and [ ] You can enter a 6-digit number using the characters [0-9], [A-F] and [ - ]. Space character is displayed as [ _ ]. Operation Display 1. Call up menu + When releasing, the first function is displayed [ ] Hold F-key while TARE –key is pressed until [ ] is displayed. see chapter 8. 2. Activate function or 3. Display ID nr. The last stored number appears 4. Input ID nr.
Select the number to be changed (the active position flashes): or Changing the numerical value or 5. Save: Your setting will be stored and the next menu item will appear. 6.
14. Data output The regular equipment of the balance includes an RS 232C interface and a printer interface. 14.1. RS 232C interface The RS 232C interface allows a bi-directional data exchange from the balance to external devices. This data exchange is asynchronous using ASCII - Code. Pin allocation of balance output plug: Pin nr.
14.2. Printer interface (unidirectional data exchange) Pin allocation of balance output plug: Pin nr. Signal Input/Output Function 1 EXT.TARE Input External tare function 2 - 3 - 4 TXD Output Transmit data 5 GND - Signal ground 6 - - 7 - - 8 - - 7 8 6 1 3 4 5 2 14.3. Description of interface The selection of a certain operating mode allows you to set the output format, the output control, the transmission speed and the parity bit.
14.4. Data output 14.4.1. Format for data transmission Any of the data formats below may be set by selecting the relevant function on the balance ( see menu overview chpt. 7.
14.4.4. Units U 1, U 2 = 2 words as ASCII-Codes U1 (SP) K C P (SP) U2 G G T C % Code 20H 47H 4BH 47H 43H 54H 50H 43H 20H 25H 14.4.5. S 1 = 1 word 86 Meaning Gram Kilogram Karat Qty.
14.4.6. S 2 = 1 word Data status S2 Code Meaning S 53 H Data stabilized * U 55 H Data not stabilized (fluctuating) * E 45 H Data error, all data apart from S 2 unreliable Balance indicating error (o-Err, u-Err) sp 20 H No special status 14.4.7. Interval data output When an interval output is started or stopped, this will result in the output of a header or footer line.
14.5. Remote control instructions 88 C1 C2 Code Meaning 0 0 4FH 30H No data output 0 1 4FH 31H Continuous data output 0 2 4FH 32H Continuous data output stable weighing values 0 3 4FH 33H Output for stable and instable weighing values after pressing PRINT key 0 4 4FH 34H Output for stable weighing value after previous relief of balance 0 5 4FH 35H One output for stable weighing value. No output for stable weighing values.
15. Service, maintenance, disposal 15.1. Cleaning Before cleaning, please disconnect the appliance from the operating voltage. Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. The weighing terminal has a pressure compensation device. This is underneath the terminal and made of a glued on membrane. When cleaning please ensure that the membrane is not damaged or soiled. 15.2.
Fault Possible cause The displayed weight does not glow. The balance is not switched on. The mains supply connection has been interrupted (mains cable not plugged in/faulty). Power supply interrupted. The displayed weight is permanently changing Draught/air movement Table/floor vibrations The weighing plate is in contact with foreign matter.