User manual
12        PCD-BA-e-1211 
The empty battery is indicated on the display by “LO“. Press ON/OFF-key and 
replace the battery immediately. 
If the balance is not used for a longer time, take out the batteries  and store them 
separately. Leaking battery liquid could damage the balance. 
If there exists an optional rechargeable battery, it has to be connected in the battery 
compartment via a separate plug-in socket. Now the mains adapter delivered with 
the rechargeable battery must be applied. 
5.5  Connection of peripheral devices 
Before connecting or disconnecting of additional devices (printer, PC) to the data 
interface, always disconnect the balance from the power supply. 
With your balance, only use accessories and peripheral devices by KERN, as they 
are ideally tuned to your balance. 
5.6  Initial Commissioning 
In order to obtain exact results with the electronic balances, the balance must have 
reached the operating temperature (see warming up time chap. 1. During this 
warming up time the balance must be connected to the power supply (mains, 
accumulator or battery). 
The accuracy of the balance depends on the local acceleration of gravity. 
Strictly observe hints in chapter Adjustment. 
5.7  Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated - in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). 
This adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation. 
Connection of rechargeable battery 
Connection of battery 