User manual
PCB-BA-e-1315 
16 
  Weigh the material, the net weight will be indicated. 
The taring process can be repeated any number of times, 
e.g. when adding several components for a mixture 
(adding). The limit is reached when the whole weighing 
range is exhausted. 
The weight of the weighing container will be displayed as a 
minus number after removing the weighing container. 
The tare weight is saved until it is deleted. 
Delete tare 
TARE
 Unload the balance and press the TARE button, the zero 
display appears. 
PRE-TARE-
Function 
MODE
CAL
Using this function the weight of a tare vessel is stored. Even 
after turning off/on the weighing balance will continue working 
with the saved tare value. 
 In weighing mode put tare vessel on the weighing plate 
 Press repeatedly the MODE key until „PtArE“ flashing 
appears. 
 Use SET key to store the current weight on the weighing 
plate as a PRE-TARE value. 
Delete PRE-TARE 
value 
MODE
CAL
 Remove all loads from the balance and press repeatedly 
the MODE key until „PtArE“ flashing appears. 
 Use the SET key to confirm. The PRE-TARE value is 
deleted, the zero display appears. 