User manual
PCB-BA-e-1315 
33 
Baud rate 
The baud rate defines the transfer speed vie the interface, 1 Baud = 
1 Bit/second. 
  In weighing mode keep the PRINT key pressed until [Unit] 
appears. 
  Press the MODE key several times until „bAUd“ is displayed. 
  Acknowledge using SET key, the current setting is displayed. 
  Use MODE key select the desired settings 
9600  4800  2400  1200  19200 
  Use the SET key to confirm selection. The balance returns to 
weighing mode. 
10  Data output RS 232 C 
10.1  Technical Data 
  8-bit ASCII Code 
  1 start bit, 8 data bits, 1 stop bit, no parity bit 
  Baud rate selectable at 1200, 2400, 4800 , 9600 and 19200 baud 
  Miniature plug-in necessary (9 pole D-Sub) 
  For  operation  with  interface  faultless  operation  is only  ensured  with  the  correct 
KERN – interface cable (max. 2m) 
10.2  Pin allocation of balance output bushing: 
Front view: 
5     1 
 9 6  
Pin 2:  Transmit data 
Pin 3:  Receive data 
Pin 5:  Signal ground 