5  Determining density of liquids 
5.1  Volume determination of the sinker 
  Prepare balance as described in chapter 2 "Installation of density determination set". 
  Fill water into the container. Filling height should be approx. ¾ of the capacity. Heat 
sample liquid until temperature is constant. 
  Keep the sinker ready 
  Switch-on balance, repeatedly press the 
UNIT
key until the balance is in weighing mode.. 
  Place sinker in the upper sample dish. Wait for 
stability display to appear; write down the 
displayed weighing value. 
  Place sinker in the lower sample dish. Wait for 
stability display to appear; write down the 
displayed weighing value. 