User manual
6 PBS-A03/A04-BA-e-1213
2 Installing the density determination set
• If necessary, carry out necessary adjustment before installation of the density
set.
• When the density set is installed, correct adjustment is not possible.
• For reasons of adjustment, take away the density set and place the standard
weighing plate.
• The following pictures show the density set KERN PBS-A03 at a balance with
big weighing plate. The density set KERN PBS-A04 must be installed at the
same way.
1. Switch off balance and separate it from the power supply.
2. Remove standard weighing plate.
3. Remove carrier of the standard
weighing plate and replace by
weighing tray carrier of the density
set.
4. Put balance tray holder according to
fig. into the four balance tray carriers.
Thereby observe the right position,
the openings on the upper side must
point forwards.