User manual
PBS-A03/A04-BA-e-1213 13
5 Determining density of liquids
5.1 Volume determination of the sinker
Prepare balance as described in chapter 2 "Installation of density determination set".
Fill water into the container. Filling height should be approx. ¾ of the capacity. Heat
sample liquid until temperature is constant.
Keep the sinker ready
Switch-on balance, repeatedly press the
UNIT
key until the balance is in weighing mode..
Place sinker in the upper sample dish. Wait for
stability display to appear; write down the
displayed weighing value.
Place sinker in the lower sample dish. Wait for
stability display to appear; write down the
displayed weighing value.