User manual
PBS/PBJ-BA-e-1314      62 
16.3.3  Specific measurement of liquid weight 
In the case of specific measurement of liquid weight, the weight of reference solid 
with known volume is measured in air and the liquid to be tested. The specific liquid 
weight is calculated on the basis of those both values. 
The display unit for the specific liquid weight is „d”.  
Density determination becomes even easier when an optional set for density 
determination is applied. For further information please refer to the operating 
instructions enclosed with the set for density determination. 
Select the menu element 61 in the menu overview. 
•  Enter the sinker density value. 
•  Remove the cover of the hook for underfloor weighing in the balance bottom. 
•  Attach the sinker on the hook and immerse it in the container filled with the 
liquid to be tested. 
•  Restore the weighing mode with the 
ESC
ON/OFF
 button and press 
UNIT
 button 
repeatedly until the „d” symbol is displayed. 
•  Place the sinker on the balance plate. 
•  When the   stabilisation display is illuminated, press 
MENU
CAL
 button.  (If 
necessary, the „dSP oL” symbol will be displayed, which does not indicate any 
disturbance.) 
•  Place the reference weight on the scale pan and immerse it in the tested 
liquid. The specific weight of the tested liquid will be displayed. 
Notes: 
•  4 positions after the point is displayed for the specific weight. If it is not 
possible to stabilize the balance with all four decimal positions, use the 1d/10d 
switch over function (refer to chapter 14.1. gelöscht).  
•  If the reference weight lies on the weighing pan immersed in the liquid, make 
sure that the whole weight is immersed in the liquid. 
16.4  Extreme value recording 
(models with non-verifiable setting only) 
The “extreme value” stands for the highest or lowest value indicated after a value 
alteration of more than fivefold the null region. 
To determine the extreme value, choose menu item 49. 