16.3  Density determination 
16.3.1  Specific measurement of solid weight 
In the case of specific measurement of solid weights, the sample (solid) weight is 
measured in air as well as in the liquid with known density and the sample density is 
calculated on the basis of that results. The ▼ symbol presents the solid density on 
this balance. Density determination with the help of the flush-mounted platform 
facility is described in the following. 
Density determination becomes even easier when an optional set for density 
determination is applied. For further information please refer to the operating 
instructions enclosed with the set for density determination. 
Select the menu element 60 in the menu overview. 
•  From the below table (chapter 16.3.2) enter the density value (g/cm³) of the 
liquid (water, alcohol etc.) in which the sample is immersed.  (Entering 
numerical values, refer to chapter 11.4, setting decimal point, refer to chapter 
11.7). To write off the setting, zero the value. 
•  Remove the cover of the hook for underfloor weighing in the balance bottom. 
•  Attach the scale pan to be hung on the hook and immerse it in the container 
filled with the liquid with known density. 
•  Restore the weighing mode with the 
ESC
ON/OFF
 button and repeatedly press the 
UNIT
 key until the „▼d” (inverted triangle and „d”) symbol is displayed. 
•   Press the 
TARE /
 button 
•  Place the sample on the weighing plate. (If necessary, the „dSP oL” symbol 
will be displayed, which does not indicate any disturbance.) 
•  When the   stabilisation display is illuminated, press 
MENU
CAL
 button.  
•  Place the sample on the scale pan to be hung. The density of the sample is 
shown. 
Notes: 
•  4 positions after the point is displayed for the specific weight.  If it is not 
possible to stabilize the balance with all four positions after the point, use the 
1d/10d switch over function (refer to chapter 12.8).  
•  If the sample lies on the scale pan immersed in the liquid, make sure that the 
whole sample is immersed in the liquid. 
•  The balance is not zeroed again if the 
TARE /
 key is pressed within this 
function. 