User manual
35        PBS/PBJ-BA-e-1314 
8.3   Automatic adjustment by PSC (Perfect Self Calibration), only models PBJ 
The PSC function determines the ambient temperature of the scale at any time. If 
the tolerance groups are exceed upwards or downwards, this is signalled and the 
necessary adjustment is carried out fully automatically. This will ensure that the scale 
is ready at any time. 
Activate function: 
  Press   3 times in weighing mode. 
Menu group 1 „Adjustment“ is displayed, the indicator   
flashes. 
  Acknowledge by  , the current setting flashes. 
   Press   several times until „A“ flashes. 
  Acknowledge with  . 
   is used to change-over among the following 
settings. 
„PSC on“ (menu element 5) = function activated 
„PSC off“ (menu element 6) = function deactivated 
The current setting is marked by the stability display 
( ). 
 
  Acknowledge selection by   . 
 
  Press   repeatedly or for 3 s and the scale returns 
into weighing mode. 