79        PBS/PBJ-BA-e-1314 
Error message: 
Displayed error 
code 
Explication 
Remedy 
Disturbances in weight loading  
mechanical elements.  
Check transport screws. 
Weight on the scale pan is unstable 
during calibration. 
Large zero point drift during 
calibration. 
Empty the scale pan. 
Large drift during the PCAL function. 
Use the correct weight. 
Large drift during calibration of the 
measuring range. 
Use the correct weight. 
Calibration weight is incorrect.  
Use the correct weight. 
Disturbances in the balance (the 
balance is stopped when this display 
appears). 
∗  
Received command code is 
incorrect. 
Check limiters etc. 
Total number of the displayed unit is 
longer than 7 positions. 
Reduce load. 
Disturbance in the balance. 
∗
The power voltage is incorrect. 
Check the power voltage . 