PBS/PBJ-BA-e-1314      66 
16.10 Automatic saving and zeroing 
This function is used for weighing large number of single samples.  If the function is 
activated, the   automatic saving and zeroing symbol is illuminated. 
Select the menu element 52 in the menu overview. 
•  Place the container to be weighed on the balance and press the 
TARE /
button when the automatic saving and zeroing function is in the stand-by 
mode.  (The   automatic saving and zeroing symbol and   mode 
symbol are illuminated.) 
  The balance will be zeroed. 
•  Press the 
PRINT
 button. The stand-by symbol disappears and measurements 
within the automatic saving and zeroing function can be started. 
•  Place the first sample on the scale pan. Each time when the   stabilisation 
symbol is illuminated and the value corresponding to five times the zero range 
or higher is illuminated or when the 
PRINT
 button is pressed, the illuminating 
value output and balance zeroing is carried out.  
•  Weighing process of the next sample is carried out without the necessity to 
press the 
ESC
ON/OFF
 button.  
•  Press the 
ESC
ON/OFF
 button. The balance returns to the stand-by mode of the 
automatic saving and zeroing function, the total weight placed on the scale 
pan without the packaging weight is displayed. To print this value, press the 
PRINT
 button. 
Notes: 
•  If the stabilisation symbol is illuminated and the display value lies within the 
zero range, zeroing is carried out automatically. 
•  Pressing the 
PRINT
 button when the display value lies below five times the zero 
range is followed by zeroing after data output.  (Manual loading) 
•  Pressing the 
ESC
ON/OFF
 button when the automatic saving and zeroing function is in 
stand-by mode is also followed by setting mains supply in stand-by mode. 