User manual
53        PBS/PBJ-BA-e-1314 
13  Functions for adaptation to ambient conditions 
13.1  Stability and reaction (average value) 
Exists the possibility to tune the stability of the display and the degree of reaction of 
the balance to the requirements of certain applications or the installation 
environment. There is a possibility of selection among five operating modes. Please 
note that in general slowing down reaction times result in higher stability of the set 
data handling, while speeding up reaction times have an influence on the stability 
deterioration.  However, the balances of PBS/PBJ type are designed to ensure both 
properties i.e. fast reaction time and high stability. 
13.1.1  Automatic mode 
Select the menu element 22: 
The balance automatically and dynamically undertakes to optimally determine the 
average value during observation of load data.   If there are not any special 
circumstances, this setting should always be used. 
13.1.2  Filling mode 
Select the menu element 23: 
This mode is suitable for weighing invariable liquid volumes. It is very susceptible to 
wind and vibration. 
(When the filling mode is active, the 
MODE
 button makes it possible to switch over  
between 3 stability settings.) 
13.1.3  Standard mode 
Select the menu element 24: 
This mode is suitable for weighing in normal environment. The determination of the 
average value is fixed and is not subject to adaptation as in the automatic mode 
which is performed dynamically.  
13.1.4  Antivibration mode 
Select the menu element 25: 
This mode should be used if the balance is placed at a location where there are 
strong vibrations, and its readings fluctuate in automatic mode.  
The balance reaction is deteriorated in a result of small changes of weight. 
13.1.5  Antiwind mode 
Select the menu element 26: 
This mode should be used if the balance is placed at a location where it is exposed 
to air currents which cause fluctuations of its readings in automatic mode.  
This reaction is deteriorated even more than in the antivibration mode, but the 
weighing process is more stable. 