User manual
PBS/PBJ-BA-e-1314      38 
8.5  ISO/GLP log 
Quality assurance systems require printouts of weighing results as well as of correct 
adjustment of the balance stating date and time and balance identification. The 
easiest way is to have a printer connected. 
Make sure that the communication parameter of balance and printer are the 
same. 
Communication parameter, see chap. 17.4 
8.5.1  Setting of adjustment protocol and scale identification number 
  Actuate   in weighing mode until „S“ flashes. 
  Acknowledge with  . 
  Press   several times until „t“ flashes. 
  Acknowledge with  . 
  Press   
   is used to change-over among the following 
settings. 
„GLP on“ (menu element 68) = function activated 
„GLP off“ (menu element 69) = function deactivated 
The current setting is marked by the stability display ( ). 
 