User manual
PBS/PBJ-BA-e-1314      36 
•  The flashing weight symbol   signals that automatic adjustment is 
imminent
.
•  In order to avoid that adjustment is started possibly during a series of 
measurements, actuate   as soon as the weight symbol flashes. This 
will interrupt automatic adjustment. 
•  If the PSC function has not been activated, the operator has to adjust by 
means of an internal adjustment weight (chapter 8.1.1) as soon as the 
weight symbol   flashes. 
8.4 Automatic adjustment by Clock-CAL (model PB only) 
With the help of its internal adjusting weight and integrated clock the balance can be 
set to carry out automatic adjustment at set times (up to three times daily, „ACALt1“, 
„ACALt2“ und „ACALt3“). Clock-CAL is a very convenient function, when calibration 
logs are desired to be made for regular calibrations, or when wishing span 
calibrations during break times to avoid interruption of measurement work. 
In order to execute Clock-Cal the following conditions must be met at the defined 
time. If the conditions are not met within one minute, adjustment is skipped. 
o  The scale must be in weighing or standby mode. 
o  The stability display must be indicated. 
o  The load at the weighing plate must be near zero. 
o  No other adjustment process must be started. 
o  The weight symbol   blinks for about two minutes as notification of span 
calibration before it begins. 
o  In order to avoid that adjustment is started possibly during a series of 
measurements, actuate   as soon as the weight symbol flashes. This 
will interrupt automatic adjustment. 
o  If all three times ate set to „00:00“ the function is switched off. 