PBS/PBJ-BA-e-1314      32 
8.2.1  Adjustment test with external weight 
•  Condition: Menu setting „E-tESt“ / Menu element 4 
•  At verified balances, the adjustment test is locked by a switch (except 
accuracy class I). To disable the access lock, destroy the seal and actuate 
the adjustment switch. Position of the adjustment switch see chap. 9 
Attention: 
After destruction of the seal the balance must be re-verified by an authorised 
agency and a new verification wire/seal mark fitted before it can be reused 
for applications subject to verification. 
  In weighing mode press  . E-tESt is displayed. 
If „E-tESt“ is not displayed, press   to return into weighing mode and 
activate menu element 4, see menu overview. 
  Press  , checking is started. The displayed 
weight value of the recommended adjustment weight 
(see chapter 1) flashes. 
If the weight value shall be changed, press   , the active digit flashes. Carry out the 
desired setting using navigation buttons (see chapter 3.1.1 „Numeric input“). 
  Carefully place adjusting weight in the centre of the 
weighing plate ,press  .  
  Wait until the zero display flashes. 