8.2  Adjustment test 
Activate function: 
  Press   3 times in weighing mode. 
Menu group 1 „Adjustment“ is displayed, the indicator   
flashes. 
  Acknowledge by  , the current setting flashes. 
Adjustment with internal weight „I-CAL“ (only 
models PBJ, see chap. 8.1.1), Menu element 1 
Adjustment test with internal weight „I-tESt“ (only 
models PBJ, see chap. 8.2.2), Menu element 2 
Adjustment with external weight „E-CAL“  
(see chap. 8.1.2), Menu element 3 
Adjustment test with external weight „E-tESt“  
(see chap. 8.2.1), Menu element 4 
  Acknowledge by  . 
  Press   repeatedly or for 3 s and the scale returns 
into weighing mode 