User manual
PBS/PBJ-BA-e-1314      28 
If „I-CAL“ is not displayed, press   to return into weighing mode and 
activate menu element 1, see menu overview. 
2.   Press the 
TARE /
 button and adjustment will take place 
automatically. 
3.  After successful adjustment the balance automatically 
returns to weighing mode. 
In case of an adjustment error (e.g. objects on the 
weighing plate) the display will show an error message, 
repeat adjustment. 
When an optional printer is connected and the GLP 
function is connected, the adjustment log will be edited, 
see chap. 8.5 
 
 
 
 
Printout example (KERN YKB-01N): 
 - - - - - - - - - - - - - - - - - 
 CAL –INTERNAL 
 - - - - - - - - - - - - - - - - - 
Adjustment mode 
 KERN 
Company 
TYPE 
PBJ4200-2M 
Model 
SN 
WBxxxxxxxxx 
Serial no. 
ID 
1234 
Balance identification no. (see chap. 
8.5.1) 
DATE 
27-01-2011 
TIME 
11.54.53 
REF 
4000.00g 
Used adjustment weight 
BFR 
4003.97g 
Before adjustment 
AFT 
4000.00g 
After adjustment 
 -COMPLETE 
 -SIGNATURE- 
prepared by 
 - - - - - - - - - - - - - - - - - 