User manual
PBS/PBJ-BA-e-1314      76 
17.4.3  Format 
Specify the format of data output by the balance. 
For the standard format of the KERN electronic balances  
•  Select menu element 77 . 
For the old data output format of the KERN electronic balances: 
•  Select menu element 78 . 
17.4.4  Speed of communication 
Specify the transmission speed (300, 600, 1200, 2400, 4800, 9600, 19200 or 
38400 bps). 
The value presented as „b-xxx” gives the number of bps (bits/second).  Baud-rate 
and bps are the same value. 
Select one of the menu elements from 81 to 88. 
17.4.5  Parity / Bit length 
Select the parity and bit length. 
No parity, 8-bit length: 
Select the menu element 89 
Odd parity, 7-bit length: 
Select the menu element 90 
Straight parity, 7-bit length: 
Select the menu element 91 
17.4.6   Stop Bits 
Select the number of stop bits. 
Stop bit 1: 
Select the menu element 92 
Stop bit 2: 
Select the menu element 93 
17.4.7   Limiter 
The „limiter” is used to separate single data or commands from each other.  The 
limiter is to be set in the following way: 
Setting to CR(0DH): 
Select the menu element 94 
Setting to LF(0AH): 
Select the menu element 95 
Setting to CR+LF(0D0AH): 
Select the menu element 96 