63        PBS/PBJ-BA-e-1314 
16.5  Auto Print function 
(models with non-verifiable setting only) 
Using the Auto Print function ensures automatic data printing without necessity to 
press the 
PRINT
 button for each single measurement.  If the function is activated, the 
 (Auto-Print) symbol is displayed. 
It is possible to select from six types of automatic printing. Information concerning 
setting the zero range can be found in chapter 11.7 
Printing whilst loading:  
Select the menu element 42 in the menu overview. 
Place or remove the sample when the displayed value lies within the zero range. The 
data will be output automatically when the   stabilisation symbol is illuminated and 
the displayed positive value exceeds 5 times the zero range value.  The next data 
output will occur when the display is brought to the value within the zero range by 
removing the sample or pressing the 
TARE /
 button. 
Printing whilst loading and unloading:   
Select the menu element 43 in the menu overview. 
Place or remove the sample when the displayed value lies within the zero range. The 
data will be output automatically when the   stabilisation symbol is illuminated and 
the displayed positive or negative value exceeds 5 times the zero range value.  The 
next data output will occur when the display is brought to the value within the zero 
range by removing the sample or pressing the 
TARE /
 button. 
Printing whilst loading and at zero:   
Select the menu element 44 in the menu overview. 
Place or remove the sample when the displayed value lies within the zero range. The 
data will be output automatically when the   stabilisation symbol is illuminated and 
the displayed positive value exceeds 5 times the zero range value.  Remove the 
sample or press the 
TARE /
 button. The data will be output again when the displayed 
value lies within the zero range and the   stabilisation symbol is illuminated. 
Printing whilst loading, unloading and at zero:  
Select the menu element 45 in the menu overview. 
Place or remove the sample when the displayed value lies within the zero range. The 
data will be output automatically when the   stabilisation symbol is illuminated and 
the displayed positive or negative value exceeds 5 times the zero range value.  
Remove the sample or press the 
TARE /
 button. The data will be output again when 
the displayed value lies within the zero range and the   stabilisation symbol is 
illuminated. 